Performance review of a slide scanner based automated counting system for PADC radon detectors

被引:8
作者
Ibrahimi, Z. -F. [1 ]
Miles, J. C. H. [1 ]
机构
[1] Hlth Protect Agcy, Radiat Protect Div, Ctr Radiat Chem & Environm Hazards, Didcot OX11 0RQ, Oxon, England
关键词
SSNTD; PADC; CR-39; etched track detector; HPA; NRPB; radon; radon-222; automatic counting; intercomparison; Blind Test; performance;
D O I
10.1016/j.radmeas.2008.03.023
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The HPA operates a radon measuring service determining the radon concentration (Bq m(-3)) in enclosed areas using passive PADC (CR-39) detectors. Since 2003 over 116,000 radon detectors have been analysed in the laboratory using a Nikon LS2000 slide scanner. Digital images of the etched PADC are analysed by in-house etched-track and area counting software. The slide scanner method is briefly described along with its performance across the dynamic range of the detectors. Blind Tests, validation and other quality assurance performance results are reported and discussed. Crown Copyright (C) 2008 Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:S395 / S400
页数:6
相关论文
共 9 条
[1]  
[Anonymous], NRPBM1140
[2]  
BECK TR, 2007, RAD PROTECT DOSIMETR
[3]   Ageing and fading of alpha particle tracks in CR-39 exposed to air [J].
Hardcastle, GD ;
Miles, JCH .
RADIATION PROTECTION DOSIMETRY, 1996, 67 (04) :295-298
[4]  
Hardcastle GD, 1996, NRPBR283
[5]  
HOWARTH CB, 2007, HPARPD028
[6]  
MILES J, 1992, APPL RADIAT ISOTOPES, V43, P87, DOI 10.1016/0883-2889(92)90081-O
[7]   Practical procedures for a radon etched track dosimetry service [J].
Miles, JCH ;
Kendall, GM ;
Ibrahimi, ZF ;
Howarth, CB .
JOURNAL OF RADIOLOGICAL PROTECTION, 2004, 24 (02) :165-171
[8]  
MILES JCH, 1997, RADON MEASUREMENTS E
[9]   Development of a reader for trace etch detectors based on a commercially available slide scanner [J].
Steele, JD ;
Bhakta, JR ;
Tanner, RJ ;
Bartlett, DT .
RADIATION MEASUREMENTS, 1999, 31 (1-6) :179-184