Leakage of fluid past the tracheal tube cuff in a benchtop model

被引:100
作者
Young, PJ [1 ]
Rollinson, M [1 ]
Downward, G [1 ]
Henderson, S [1 ]
机构
[1] CHRISTCHURCH HOSP,INTENS CARE UNIT,CHRISTCHURCH,NEW ZEALAND
关键词
equipment; tubes tracheal; model; trachea; complications; pneumonia; aspiration; intubation tracheal;
D O I
10.1093/bja/78.5.557
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have assessed a range of high volume, low pressure (HVLP) cuffed tracheal tubes in a benchtop model, for leakage of fluid from above the cuff to the model trachea below, during various ventilatory modes. Rapid leakage occurred in the model during all modes of ventilation, unless tracheal pressure was greater than the height of fluid in the column above the cuff. This leakage occurred preferentially down longitudinal folds that occur in the HVLP cuff wall. This model suggests that, if a longitudinal fold within the cuff wall is patent, then the possibility exists of subglottic to tracheal leakage.
引用
收藏
页码:557 / 562
页数:6
相关论文
共 7 条
[1]   PREVENTION OF HOSPITAL-ACQUIRED PNEUMONIA - MEASURING EFFECT IN OUNCES, POUNDS, AND TONS [J].
CRAVEN, DE .
ANNALS OF INTERNAL MEDICINE, 1995, 122 (03) :229-231
[2]   THE PATHOGENESIS OF VENTILATOR-ASSOCIATED PNEUMONIA .1. MECHANISMS OF BACTERIAL TRANSCOLONIZATION AND AIRWAY INOCULATION [J].
ESTES, RJ ;
MEDURI, GU .
INTENSIVE CARE MEDICINE, 1995, 21 (04) :365-383
[3]   PREVENTION OF NOSOCOMIAL PNEUMONIA IN INTUBATED PATIENTS - RESPECTIVE ROLE OF MECHANICAL SUBGLOTTIC SECRETIONS DRAINAGE AND STRESS-ULCER PROPHYLAXIS [J].
MAHUL, P ;
AUBOYER, C ;
JOSPE, R ;
ROS, A ;
GUERIN, C ;
ELKHOURI, Z ;
GALLIEZ, M ;
DUMONT, A ;
GAUDIN, O .
INTENSIVE CARE MEDICINE, 1992, 18 (01) :20-25
[4]  
MEHTA S, 1984, ANN ROY COLL SURG, V66, P356
[5]   FAILURE OF A HIGH-COMPLIANCE LOW-PRESSURE CUFF TO PREVENT ASPIRATION [J].
PAVLIN, EG ;
VANNIMWEGEN, D ;
HORNBEIN, TF .
ANESTHESIOLOGY, 1975, 42 (02) :216-219
[6]   ENDOTRACHEAL CUFF PRESSURE AND TRACHEAL MUCOSAL BLOOD-FLOW - ENDOSCOPIC STUDY OF EFFECTS OF 4 LARGE VOLUME CUFFS [J].
SEEGOBIN, RD ;
VANHASSELT, GL .
BRITISH MEDICAL JOURNAL, 1984, 288 (6422) :965-968
[7]   ASPIRATION BEYOND ENDOTRACHEAL CUFFS [J].
SEEGOBIN, RD ;
VANHASSELT, GL .
CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1986, 33 (03) :273-279