Characterization of a magnetic bearing system and fluid properties for a continuous flow ventricular assist device

被引:13
作者
Baloh, MJ [1 ]
Allaire, PE
Hilton, EF
Wei, NH
Olsen, SB
Bearnson, SB
Khanwilkar, PS
机构
[1] Univ Virginia, Dept Elect Engn, Charlottesville, VA 22903 USA
[2] Univ Virginia, Dept Mech Aerosp & Nucl Engn, Charlottesville, VA 22903 USA
[3] Univ Utah, Artificial Heart Lab, Salt Lake City, UT USA
[4] MedQuest Prod Inc, Salt Lake City, UT USA
关键词
ventricular assist device; continuous flow blood pump; pump performance;
D O I
10.1046/j.1525-1594.1999.06425.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This article presents the performance test results of the CFVAD3 continuous flow blood pump in an artificial human circulation system. The CFVAD3 utilizes magnetic bearings that support a thin pancake impeller, the shape of which allows for a very compact pump whose total axial length is less than 5 cm with a radial length of about 10 cm. This gives a total volume of about 275 cc. The impeller itself has 4 vanes with a designed operating point of 6 L/min at 100 mm Hg of differential pressure and 2,000 rpm. The advantages of magnetic bearings, such as large clearance spaces and no mechanical wear, are elaborated upon. Furthermore, bearing modal parameters such as load capacity and current gains are described. These parameters in conjunction with the operating conditions during testing are then used to estimate the fluid forces, stiffness, and damping properties while pumping. Knowledge of these parameters is desirable because of their effects on pump behavior. In addition, a better plant model will allow more robust control algorithms to be devised that can boost pump performance and reliability.
引用
收藏
页码:792 / 796
页数:5
相关论文
共 9 条
[1]  
Allaire P, 1998, ARTIF ORGANS, V22, P475
[2]  
ALLAIRE PE, 1996, TRIBOL T, V39
[3]   Development of a prototype magnetically suspended rotor ventricular assist device [J].
Bearnson, GB ;
Maslen, EH ;
Olsen, DB ;
Allaire, PE ;
Khanwilkar, PS ;
Long, JW ;
Kim, HC .
ASAIO JOURNAL, 1996, 42 (04) :275-281
[4]  
HILTON EF, 1998, AM SOC ARTIF INTERN
[5]  
HOGNESS JR, 1991, ARTIFICIAL HEART PRO, P65
[6]   SYSTEM CONSIDERATIONS FAVORING ROTARY ARTIFICIAL HEARTS WITH BLOOD-IMMERSED BEARINGS [J].
JARVIK, RK .
ARTIFICIAL ORGANS, 1995, 19 (07) :565-570
[7]  
MASLEN EH, 1996, IEEE T INSTRUMENT ME
[8]   THE NEXT-GENERATION BAYLOR C-GYRO PUMP - ANTITHROMBOGENIC FREE IMPELLER DESIGN FOR LONG-TERM CENTRIFUGAL VAD [J].
OHARA, Y ;
NOSE, Y .
ARTIFICIAL ORGANS, 1994, 18 (03) :238-243
[9]  
WATERS T, 1998, IN PRESS J AM SOC AR