A cryogenic current comparator bridge for resistance measurements at currents of up to 100 A

被引:10
作者
Williams, JM [1 ]
Kleinschmidt, P [1 ]
机构
[1] Natl Phys Lab, Ctr Electromagnet Metrol, Teddington TW11 0LW, Middx, England
关键词
cryogenic current comparator; resistance;
D O I
10.1109/19.769604
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The use of cryogenic current comparators at NPL has been limited so far to currents of 100 mA, A new current comparator with a maximum current capacity of 100 A has been developed and forms the basis of a resistance ratio bridge for low value resistors with a relative uncertainty of approximately 10(-7) or better.
引用
收藏
页码:375 / 378
页数:4
相关论文
共 4 条
[1]   IRONLESS CRYOGENIC CURRENT COMPARATORS FOR AC-APPLICATIONS AND DC-APPLICATIONS [J].
GROHMANN, K ;
HAHLBOHM, HD ;
LUBBIG, H ;
RAMIN, H .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1974, IM23 (04) :261-263
[2]   CRYO CURRENT COMPARATOR AS A CALCULABLE DC RATIO STANDARD [J].
GROHMANN, K ;
HAHLBOHM, HD ;
HECHTFISCHER, D .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1979, 28 (03) :205-211
[3]   AN AUTOMATED CRYOGENIC CURRENT COMPARATOR RESISTANCE RATIO BRIDGE [J].
WILLIAMS, JM ;
HARTLAND, A .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1991, 40 (02) :267-270
[4]  
WILLIAMS JM, 1993, 6 BRIT EL MEAS C TED