Helical CT imaging performance of a new multislice scanner

被引:2
作者
Hu, H [1 ]
He, HD [1 ]
Foley, WD [1 ]
Fox, SH [1 ]
机构
[1] GE Co, Med Syst, Milwaukee, WI 53201 USA
来源
MEDICAL IMAGING 1999: IMAGE PROCESSING, PTS 1 AND 2 | 1999年 / 3661卷
关键词
helical CT; multi-slice CT; imaging performance; comparative studies;
D O I
10.1117/12.348600
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The principles and operations of 4-slice helical CT are discussed. The SSP, image noise and artifacts of a 4-slice scanner are measured by phantom scans and compared with theoretical predictions and the measurements of single-slice CT. The evaluation of these physical attributes for all helical imaging modes of 4-slice scanner is summarized in an operation chart. The preliminary studies indicate that comparing with single-slice helical CT, the volume coverage speed of 4-slice helical CT can be at least twice as fast with fully comparable image quality, or, in many cases, 3 times as fast with diagnostically comparable image quality. Two examples are given to illustrate the clinical benefit of the speed performance improvement (and operation flexibility) provided by 4-slice helical CT.