A randomized trial of 5 versus 7 French guiding catheters for transfemoral percutaneous coronary stent implantation

被引:17
作者
Buchler, Jorge R. [1 ]
Ribeiro, Expedito E. [1 ]
Falcao, Joao L. [1 ]
Martinez, Eulogio E. [1 ]
Buchler, Rica D. [1 ]
Feres, Fausto [1 ]
Maielo, Jose R. [1 ]
Haddad, Nagib [1 ]
Ramires, Jose F. [1 ]
Ellis, Stephen G. [2 ]
机构
[1] Univ Sao Paulo, Incor Heart Inst, Sao Paulo, Brazil
[2] Cleveland Clin Fdn, Cleveland, OH 44195 USA
关键词
D O I
10.1111/j.1540-8183.2007.00315.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Comparative studies between 5 French guiding catheter and others of larger size using the transfemoral approach to coronary stenting have not been described. Coronary stent implantation was performed in 90 patients in a randomized trial. The primary end-point was to compare the incidence of successful uncomplicated stent implantation per lesion with the 5F and 7F guiding catheters. Patients were excluded for excessive vessel tortuosity or anticipated need for equipment not fitting through a 5 catheter. Baseline characteristics and the use of direct stenting did not differ between the two groups. The primary success rate was 97.8% per patient in both groups and 98% per lesion in the 5 French and 97.9% in the 7 French. Guiding catheter change was necessary in 1 patient in each group to successfully complete the procedure in both groups. The amount of contrast used was 63 +/- 27.3 mL in the 5 French and 76 +/- 25 mL in the 7 French groups (P < 0.05). Vascular complications and blood transfusions occurred somewhat more frequently in the 7 French group (P = 0.058). The manual compression time after sheath removal was 5.1 +/- 2.0 min and 8.0 +/- 4.3 min, respectively, in the 5 and 7 French groups (P < 0.01). In conclusion, the 5 French guiding catheters showed a similar success rate with coronary stenting when compared to the 7 French, but the amount of contrast used and manual compression time after sheath removal, as well as the rate of vascular and bleeding complications, were reduced in the 5 French group.
引用
收藏
页码:50 / 55
页数:6
相关论文
共 13 条
[1]  
Bartorelli AL, 1999, CATHETER CARDIO INTE, V48, P279, DOI 10.1002/(SICI)1522-726X(199911)48:3<279::AID-CCD9>3.3.CO
[2]  
2-Z
[3]  
Chevalier B, 2000, EUR HEART J, V21, P288
[4]   TRANS-LUMINAL DILATATION OF CORONARY-ARTERY STENOSIS [J].
GRUNTZIG, A .
LANCET, 1978, 1 (8058) :263-263
[5]   Intracoronary stent implantation without ultrasound guidance and with replacement of conventional anticoagulation by antiplatelet therapy - 30-day clinical outcome of the French Multicenter Registry [J].
Karrillon, GJ ;
Morice, MC ;
Benveniste, E ;
Bunouf, P ;
Aubry, P ;
Cattan, S ;
Chevalier, B ;
Commeau, P ;
Cribier, A ;
Eiferman, C ;
Grollier, G ;
Guerin, Y ;
Henry, M ;
Lefevre, T ;
Livarek, B ;
Louvard, Y ;
Marco, J ;
Makowski, S ;
Monassier, JP ;
Pernes, JM ;
Rioux, P ;
Spaulding, C ;
Zemour, G .
CIRCULATION, 1996, 94 (07) :1519-1527
[6]  
Koch KT, 2001, EUR HEART J, V22, P285
[7]   Comparison of 6F with 7F and 8F guiding catheters for elective coronary angioplasty: Results of a prospective, multicenter, randomised trial [J].
Metz, D ;
Meyer, P ;
Touati, C ;
Coste, P ;
Petiteau, PY ;
Durand, P ;
Faivre, R ;
Lefevre, T ;
Elaerts, J .
AMERICAN HEART JOURNAL, 1997, 134 (01) :131-137
[8]  
Peels HOJ, 1996, CATHETER CARDIO DIAG, V38, P210, DOI 10.1002/(SICI)1097-0304(199606)38:2<210::AID-CCD21>3.0.CO
[9]  
2-L
[10]   CORONARY ANGIOPLASTY THROUGH A NEW 6 FRENCH GUIDING CATHETER [J].
RESAR, JR ;
PREWITT, KC ;
WOLFF, MR ;
BRINKER, JA .