A 4x1 GHz array acousto-optical spectrometer

被引:37
作者
Horn, J [1 ]
Siebertz, O [1 ]
Schmülling, F [1 ]
Kunz, C [1 ]
Schieder, R [1 ]
Winnewisser, G [1 ]
机构
[1] Univ Cologne, Inst Phys 1, D-50937 Cologne, Germany
关键词
radio astronomy; array systems; acousto-optical spectrometer; back-ends; radiometers; heterodyne instrumentation;
D O I
10.1023/A:1008093310165
中图分类号
P1 [天文学];
学科分类号
0704 ;
摘要
In this paper an array acousto-optical spectrometer (AOS) with four 1 GHz bands and a frequency channel spacing of 1 MHz is presented. A discussion about the requirements which have to be fulfilled by the presented instrument for applications in radio astronomy is given as well as e detailed description of the optical design. Remarkable test results have been achieved such as an Allan-Variance minimum time of about 1000 s, identical noise performance, and identical frequency calibration of all bands. The reproducibility of measurements has been demonstrated by tests with CH3CN, using an absorption cell. The question of the maximum number of bands per array AOS is discussed. Although it may seem promising to increase the number of bands per optical unit to e.g. eight or more instead of four, a strong limitation has been found due to the increasing r.f. power level on the Bragg cell which affects the baselines of the spectra.
引用
收藏
页码:17 / 38
页数:22
相关论文
共 23 条
[1]  
BOS A, 1985, SPIE P, V598, P134
[2]  
DIXON RW, 1971, PHYSICS OPTOELECTRON
[3]  
FRERICK J, 1995, THESIS U KOLN
[4]  
*GEC MARC, 1995, Y36743602 GEC MARC
[5]  
HJALMARSON A, 1993, P 2 COL ZERM S
[6]  
HORN J, 1997, THESIS U KOLN
[7]  
KLUMB M, 1994, P SOC PHOTO-OPT INS, V2268, P305, DOI 10.1117/12.185841
[8]  
Klumb M, 1995, P SOC PHOTO-OPT INS, V2583, P420, DOI 10.1117/12.228596
[9]  
Kraus J. D., 1986, RADIO ASTRONOMY
[10]   ACOUSTOOPTICAL SPECTROMETERS FOR BROAD-BAND MILLIMETER RADIOASTRONOMY AT IRAM [J].
LECACHEUX, A ;
ROSOLEN, C ;
DIERICH, P ;
PAUBERT, G .
INTERNATIONAL JOURNAL OF INFRARED AND MILLIMETER WAVES, 1993, 14 (02) :169-184