Dephased MRI

被引:36
作者
Bakker, CJG
Seppenwoolde, JH
Vincken, KL
机构
[1] Univ Utrecht, Med Ctr, Dept Radiol, NL-3584 CX Utrecht, Netherlands
[2] Univ Utrecht, Med Ctr, Image Sci Inst, NL-3584 CX Utrecht, Netherlands
关键词
dephasing; k-space filtering; background suppression; partial volume effects; susceptibility compensation;
D O I
10.1002/mrm.20733
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this work gradient dephasing is treated as a mechanism for manipulating contrast in otherwise conventional MR images. The paper provides a theoretical and experimental framework for this approach. It starts from the observation that dephasing gradients invoke a shift in k-space. From this it is inferred that the effects of in-plane and through-plane dephasing can be systematically explored in the context of any given imaging experiment by sampling k-space more widely and densely than dictated by the field of view (FOV) and the spatial resolution of the desired images. The oversampled k-space allows an ensemble of lower-resolution dephased images to be reconstructed in which the degree and direction of dephasing are determined by the off-center position of the reconstruction window. The efficacy of this approach is demonstrated for standard gradient-echo acquisitions in a phantom. The results indicate the potential of the proposed methodology for evaluating 3D image data and optimizing gradient dephasing in applications that rely on the exploitation of partial volume and susceptibility effects (e.g., tracking interventional devices and tracing magnetically labeled substances).
引用
收藏
页码:92 / 97
页数:6
相关论文
共 28 条
[1]   COMPARISON OF PHASE-DIFFERENCE AND COMPLEX-DIFFERENCE PROCESSING IN PHASE-CONTRAST MR ANGIOGRAPHY [J].
BERNSTEIN, MA ;
IKEZAKI, Y .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (06) :725-729
[2]   On the artifact of a subvoxel susceptibility deviation in spoiled gradient-echo imaging [J].
Bos, C ;
Viergever, MA ;
Bakker, CJG .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (02) :400-404
[3]   Background suppression using magnetization preparation for contrast-enhanced MR projection angiography [J].
Bos, C ;
Bakker, CJG ;
Viergever, MA .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (01) :78-87
[4]  
Bracewell R. N., 1986, FOURIER TRANSFORM IT, V31999
[5]  
CARMICHAEL DW, 2002, P 10 ANN M ISMRM HON, P296
[6]  
CARMICHAEL DW, 2005, P 13 ANN M ISMRM MIA, P2613
[7]  
CARMICHAEL DW, 2004, P 12 ANN M ISMRM KYO, P803
[8]   TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11
[9]  
Constable RT, 1999, MAGNET RESON MED, V42, P110, DOI 10.1002/(SICI)1522-2594(199907)42:1<110::AID-MRM15>3.0.CO
[10]  
2-3