Placement of a flexible endovascular stent across the femoral joint: An in vivo study in the swine model

被引:45
作者
Andrews, RT [1 ]
Venbrux, AC [1 ]
Magee, CA [1 ]
Bova, DA [1 ]
机构
[1] Johns Hopkins Univ Hosp, Cardiovasc Diagnost Lab, Baltimore, MD 21287 USA
关键词
arteries; stents and prostheses; veins;
D O I
10.1016/S1051-0443(99)70222-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To investigate the effects of joint motion on the structural integrity of periarticular stents and on the development of neointimal hyperplasia within these devices. MATERIALS AND METHODS: In four juvenile farm swine, Wallstents were implanted in the common femoral arteries and contralateral common femoral veins, centered at the point of maximal conformational change during passive hip flexion, Control stents were placed in the aortae and iliac veins. Angiography and transcatheter blood pressure measurements were obtained across each stent, with periarticular stents studied in flexion and extension, Two animals underwent repeated evaluation after 1 month, the others after 3 months. Findings were correlated with gross and histopathologic findings in the harvested stents, RESULTS: No stent fractures occurred. One femoral vein was injured during stent placement and was occluded 1 month later at follow-up, Hemodynamically significant stenoses were identified in one arterial stent and one venous stent at 3 months. The amount of neointimal hyperplasia was greater in periarticular stents than in controls and greater in animals studied at 1 month than in those studied at 3 months. The pattern of neointimal hyperplasia within mobile arteries was circumferentially asymmetric and thicker at the distal ends of the stents, Venous neointimal hyperplasia was thicker and markedly different in character than that seen in arterial stents from the same animals. CONCLUSIONS: Periarticular Wallstents and the underlying vascular anatomy remained structurally intact despite the stresses of repetitive motion during a 3-month period, Stents deployed across joints or in venous locations may be at greater risk for neointimal hyperplasia development and eventual occlusion than those deployed in immobile vessels and arteries. Neointimal hyperplasia may decrease in thickness after an initial period of exuberant development. Additional studies are necessary to determine longterm outcomes.
引用
收藏
页码:1219 / 1228
页数:10
相关论文
共 38 条
  • [1] EFFECT OF COMPLIANCE MISMATCH ON VASCULAR GRAFT PATENCY
    ABBOTT, WM
    MEGERMAN, J
    HASSON, JE
    LITALIEN, G
    WARNOCK, DF
    [J]. JOURNAL OF VASCULAR SURGERY, 1987, 5 (02) : 376 - 382
  • [2] CHANGES IN ARTERIAL-WALL COMPLIANCE AFTER ENDOVASCULAR STENTING
    BACK, M
    KOPCHOK, G
    MUELLER, M
    CAVAYE, D
    DONAYRE, C
    WHITE, RA
    [J]. JOURNAL OF VASCULAR SURGERY, 1994, 19 (05) : 905 - 911
  • [3] NEOINTIMA FORMATION AFTER VASCULAR STENT IMPLANTATION - SPATIAL AND CHRONOLOGICAL DISTRIBUTION OF SMOOTH-MUSCLE CELL-PROLIFERATION AND PHENOTYPIC MODULATION
    BAI, HZ
    MASUDA, J
    SAWA, Y
    NAKANO, S
    SHIRAKURA, R
    SHIMAZAKI, Y
    OGATA, J
    MATSUDA, H
    [J]. ARTERIOSCLEROSIS AND THROMBOSIS, 1994, 14 (11): : 1846 - 1853
  • [4] Paired comparison of vascular wall reactions to Palmaz stents, Strecker tantalum stents, and Wallstents in canine iliac and femoral arteries
    Barth, KH
    Virmani, R
    Froelich, J
    Takeda, T
    Lossef, SV
    Newsome, J
    Jones, R
    Lindisch, D
    [J]. CIRCULATION, 1996, 93 (12) : 2161 - 2169
  • [5] BERRY JL, 1995, 1995 INT MECH ENG C, P231
  • [6] COLLAPSE OF A PALMAZ STENT IN THE SUBCLAVIAN VEIN
    BJARNASON, H
    HUNTER, DW
    CRAIN, MR
    FERRAL, H
    MILTZMILLER, SE
    WEGRYN, SA
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (05) : 1123 - 1124
  • [7] MORPHOLOGIC CHARACTERISTICS OF LESION FORMATION AND TIME-COURSE OF SMOOTH-MUSCLE CELL-PROLIFERATION IN A PORCINE PROLIFERATIVE RESTENOSIS MODEL
    CARTER, AJ
    LAIRD, JR
    FARB, A
    KUFS, W
    WORTHAM, DC
    VIRMANI, R
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (05) : 1398 - 1405
  • [8] Endovascular repair of peripheral aneurysms, pseudoaneurysms, and arteriovenous fistulas
    Criado, E
    Marston, WA
    Ligush, J
    Mauro, MA
    Keagy, BA
    [J]. ANNALS OF VASCULAR SURGERY, 1997, 11 (03) : 256 - 263
  • [9] SELF-EXPANDING METALLIC STENTS FOR SMALL VESSELS - AN EXPERIMENTAL EVALUATION
    DUPRAT, G
    WRIGHT, KC
    CHARNSANGAVEJ, C
    WALLACE, S
    GIANTURCO, C
    [J]. RADIOLOGY, 1987, 162 (02) : 469 - 472
  • [10] SMALL STENT SIZE AND INTIMAL HYPERPLASIA CONTRIBUTE TO RESTENOSIS - A VOLUMETRIC INTRAVASCULAR ULTRASOUND ANALYSIS
    DUSSAILLANT, GR
    MINTZ, GS
    PICHARD, AD
    KENT, KM
    SATLER, LF
    POPMA, JJ
    WONG, SC
    LEON, MB
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) : 720 - 724