Studies on image quality, high contrast resolution and dose for the axial skeleton and limbs with a new, dedicated CT system (ISO-C-3D)

被引:53
作者
Rock, C
Kotsianos, D
Linsenmaier, U
Fischer, T
Brandl, R
Vill, F
Wirth, S
Kaltschmidt, R
Euler, E
Pfeifer, KJ
Reiser, M
机构
[1] Univ Munich, Klinikum Innenstadt, Inst Klin Radiol, D-80336 Munich, Germany
[2] Siemens AG, Medizintech, D-8520 Erlangen, Germany
[3] Univ Munich, Klinikum Innenstadt, Chirurg Klin & Poliklin, D-80336 Munich, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2002年 / 174卷 / 02期
关键词
3D-C-arm; computed tomography; high contrast; axial skeleton; joints;
D O I
10.1055/s-2002-20107
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Purpose: Evaluation of 3D-CT imaging of the axial skeleton and different joints of the lower and upper extremities with a new dedicated CT system (ISO-C-3D) based on a mobile isocentric C-arm image amplifier. Material and Methods: 27 cadaveric specimes of different joints of the lower and upper extremities and of the spinal column were examined with 3D-CT imaging (ISO-C-3D). All images were evaluated by 3 radiologists for image quality using a semiquantitative score (score value 1: poor quality; score value 4: excellent quality). In addition, dose measurements and measurements of high contrast resolution were performed in comparison to conventional and low-dose spiral CT using a high contrast phantom (Catphan, Phantom Laboratories). Results: Adequate image quality (mean score values 3 - 4) could be achieved with an applied dose comparable to low-dose CT in smaller joints such as wrist, elbow, ankle and knee. A remarkably inferior image quality resulted in imaging of the hip, lumbar and thoracic spine (mean score values 2 - 3) in spite of almost doubling the dose (dose increased by 85 percent). The image quality of shoulder examinations was insufficient (mean score value 1). Phantom studies showed a high-contrast resolution comparable to helical CT in the xy-axis (91p/cm). Conclusion: Preliminary results show, that image quality of C-arm-based CT-imaging (ISO-C-3D) seems to be adequate in smaller joints. ISO-C-3D images of the hip and axial skeleton show a decreased image quality, which does not seem to be sufficient for diagnosing subtle fractures.
引用
收藏
页码:170 / 176
页数:7
相关论文
共 17 条
[1]
Spiral CT vs incremental CT: Is spiral CT superior in imaging of the brain? [J].
Bahner, ML ;
Reith, W ;
Zuna, I ;
Engenhart-Cabillic, R ;
van Kaick, G .
EUROPEAN RADIOLOGY, 1998, 8 (03) :416-420
[2]
BERLEMANN U, 2000, AKTUEL TRAUMATOL, V30, P132
[3]
Brown J, 2000, ABSTR PAP AM CHEM S, V219, pU321
[4]
A mobile computed tomographic scanner with intraoperative and intensive care unit applications [J].
Butler, WE ;
Piaggio, CM ;
Constantinou, C ;
Niklason, L ;
Gonzalez, RG ;
Cosgrove, GR ;
Zervas, NT .
NEUROSURGERY, 1998, 42 (06) :1304-1310
[5]
EBRAHEIM NA, 1994, CLIN ORTHOP RELAT R, V307, P222
[6]
PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[7]
Freyschmidt J, 1999, ROFO-FORTSCHR RONTG, V170, P1
[8]
FRIED M, 1997, CURRENT OPINION OTOL, V511, P22
[9]
FRODEL JL, 1999, FACIAL PLASTIC SURG, V1511, P83
[10]
Gebhard F, 2000, UNFALLCHIRURG, V103, P696, DOI 10.1007/s001130050605