Tapping-mode atomic force microscopy produces faithful high-resolution images of protein surfaces

被引:195
作者
Möller, C
Allen, M
Elings, V
Engel, A
Müller, DJ
机构
[1] Univ Basel, Bioctr, ME Muller Inst Struct Biol, CH-4056 Basel, Switzerland
[2] Forschungszentrum Julich, Struct Biol IBI2, D-52425 Julich, Germany
[3] Digital Instruments, Santa Barbara, CA 93117 USA
关键词
D O I
10.1016/S0006-3495(99)76966-3
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Compared to contact-mode atomic force microscopy (CMAFM), tapping-mode atomic force microscopy (TMAFM) has the advantage of allowing imaging surfaces of macromolecules, even when they are only weakly attached to the support. In this study, TMAFM is applied to two different regular protein layers whose structures are known to great detail, the purple membrane from Halobacterium salinarum and the hexagonally packed intermediate (HPI) layer from Deinococcus radiodurans, to assess the faithfulness of high-resolution TMAFM images. Topographs exhibited a lateral resolution between 1.1 and 1.5 nm and a vertical resolution of similar to 0.1 nm. For all protein surfaces, TMAFM and CMAFM topographs were in excellent agreement. TMAFM was capable of imaging the fragile polypeptide loop connecting the transmembrane. alpha-helices E and F of bacteriorhodopsin in its native extended conformation. The standard deviation (SD) of averages calculated from TMAFM topographs exhibited an enhanced minimum (between 0.1 and 0.9 nm) that can be assigned to the higher noise of the raw data. However, the SD difference, indicating the flexibility of protein subunits, exhibited an excellent agreement between the two imaging modes. This demonstrates that the recently invented imaging-mode TMAFM has the ability to faithfully record high-resolution images and has sufficient sensitivity to contour individual peptide loops without detectable deformations.
引用
收藏
页码:1150 / 1158
页数:9
相关论文
共 72 条
[1]   COMPARATIVE-STUDY OF A REGULAR PROTEIN LAYER BY SCANNING TUNNELING MICROSCOPY AND TRANSMISSION ELECTRON-MICROSCOPY [J].
AMREIN, M ;
WANG, Z ;
GUCKENBERGER, R .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1991, 9 (02) :1276-1281
[2]  
[Anonymous], PROCEDURES SCANNING
[3]   THE MAJOR CELL-ENVELOPE PROTEIN OF MICROCOCCUS-RADIODURANS (R1) - STRUCTURAL AND CHEMICAL CHARACTERIZATION [J].
BAUMEISTER, W ;
KARRENBERG, F ;
RACHEL, R ;
ENGEL, A ;
TENHEGGELER, B ;
SAXTON, WO .
EUROPEAN JOURNAL OF BIOCHEMISTRY, 1982, 125 (03) :535-544
[4]   3-DIMENSIONAL STRUCTURE OF THE REGULAR SURFACE-LAYER (HPI LAYER) OF DEINOCOCCUS-RADIODURANS [J].
BAUMEISTER, W ;
BARTH, M ;
HEGERL, R ;
GUCKENBERGER, R ;
HAHN, M ;
SAXTON, WO .
JOURNAL OF MOLECULAR BIOLOGY, 1986, 187 (02) :241-253
[5]   MOTION AND ENZYMATIC DEGRADATION OF DNA IN THE ATOMIC-FORCE MICROSCOPE [J].
BEZANILLA, M ;
DRAKE, B ;
NUDLER, E ;
KASHLEV, M ;
HANSMA, PK ;
HANSMA, HG .
BIOPHYSICAL JOURNAL, 1994, 67 (06) :2454-2459
[6]   ATOMIC FORCE MICROSCOPE [J].
BINNIG, G ;
QUATE, CF ;
GERBER, C .
PHYSICAL REVIEW LETTERS, 1986, 56 (09) :930-933
[7]   HAS NEGATIVE STAINING STILL A PLACE IN BIOMACROMOLECULAR ELECTRON-MICROSCOPY [J].
BREMER, A ;
HENN, C ;
ENGEL, A ;
BAUMEISTER, W ;
AEBI, U .
ULTRAMICROSCOPY, 1992, 46 (1-4) :85-111
[8]   MEASURING LOCAL SURFACE-CHARGE DENSITIES IN ELECTROLYTE-SOLUTIONS WITH A SCANNING FORCE MICROSCOPE [J].
BUTT, HJ .
BIOPHYSICAL JOURNAL, 1992, 63 (02) :578-582
[9]   CALCULATION OF THERMAL NOISE IN ATOMIC-FORCE MICROSCOPY [J].
BUTT, HJ ;
JASCHKE, M .
NANOTECHNOLOGY, 1995, 6 (01) :1-7
[10]   Direct visualization of surface charge in aqueous solution [J].
Czajkowsky, DM ;
Allen, MJ ;
Elings, V ;
Shao, ZF .
ULTRAMICROSCOPY, 1998, 74 (1-2) :1-5