Smearing effects introduced by the response function of position-sensitive gas detectors in SAXS experiments

被引:19
作者
Ne, F
Gabriel, A
Kocsis, M
Zemb, T
机构
[1] ILL 20,EMBL GRENOBLE,F-38042 GRENOBLE,FRANCE
[2] EUROPEAN SYNCHROTRON RADIAT FACIL,F-38043 GRENOBLE,FRANCE
关键词
D O I
10.1107/S0021889896013568
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The response of gas-filled position-sensitive detectors consists not only of the main Gaussian response (FWHM similar to 0.1 mm) but also of a linear exponential term measurable up to 7 mm from the excitation, superimposed on an incoherent background. This effect depends on gas type and pressure and is due to fluorescence of the detection gas. The consequences of this effect on some standard experimental methods used in small-angle X-ray scattering (SAXS), i.e. Porod limit and Guinier radii determination, are investigated.
引用
收藏
页码:306 / 311
页数:6
相关论文
共 14 条
[1]   HIGH-RESOLUTION X-RAY GAS PROPORTIONAL DETECTORS WITH DELAY-LINE POSITION SENSING FOR HIGH COUNTING RATES [J].
BOIE, RA ;
FISCHER, J ;
INAGAKI, Y ;
MERRITT, FC ;
RADEKA, V ;
ROGERS, LC ;
XI, DM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 201 (01) :93-115
[2]   THE PRIMARY ELECTRON CLOUD IN XENON FOR X-RAYS IN THE 0.1 TO 10-KEV RANGE [J].
DIAS, THVT ;
SANTOS, FP ;
CONDE, CAN .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :137-139
[3]   POSITION-SENSITIVE X-RAY DETECTOR [J].
GABRIEL, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1977, 48 (10) :1303-1305
[4]  
GROH J, 1990, 90023 DESY
[5]  
Guinier A., 1939, Ann Phys, V12, P161, DOI DOI 10.1051/ANPHYS/193911120161
[6]  
Guinier G Fournet A., 1955, SMALL ANGLE SCATTERI
[7]   Effect of self-induced space charge in a high pressure position-sensitive proportional counter [J].
Ito, S ;
Tosaki, M ;
Maeda, N .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 368 (03) :738-744
[8]  
MITTELBACH P, 1964, ACTA PHS AUSTRIACA, V19, P102
[9]   CHARACTERIZATION OF AN IMAGE-PLATE DETECTOR USED FOR QUANTITATIVE SMALL-ANGLE-SCATTERING STUDIES [J].
NE, F ;
GAZEAU, D ;
LAMBARD, J ;
LESIEUR, P ;
ZEMB, T ;
GABRIEL, A .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1993, 26 (pt 6) :763-773
[10]   A POSITION-SENSITIVE DETECTOR SYSTEM FOR THE MEASUREMENT OF DIFFUSE-X-RAY SCATTERING [J].
OSBORN, JC ;
WELBERRY, TR .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1990, 23 :476-484