Microneedle Arrays Allow Lower Microbial Penetration Than Hypodermic Needles In Vitro

被引:172
作者
Donnelly, Ryan F. [1 ]
Singh, Thakur Raghu Raj [1 ]
Tunney, Michael M. [1 ]
Morrow, Desmond I. J. [1 ]
McCarron, Paul A. [2 ]
O'Mahony, Conor [3 ]
Woolfson, A. David [1 ]
机构
[1] Queens Univ Belfast, Sch Pharm, Ctr Med Biol, Belfast BT9 7BL, Antrim, North Ireland
[2] Univ Ulster, Dept Pharm & Pharmaceut Sci, Coleraine BT52 1SA, Londonderry, North Ireland
[3] Tyndall Natl Inst, Cork, Ireland
基金
英国生物技术与生命科学研究理事会; 爱尔兰科学基金会;
关键词
infection; microneedle arrays; microorganisms; stratum corneum; INTRADERMAL DELIVERY; TRANSDERMAL DELIVERY; FABRICATION; MODEL;
D O I
10.1007/s11095-009-9967-2
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
In this study we determined, for the first time, the ability of microorganisms to traverse microneedle-induced holes using two different in vitro models. When employing SilescolA (R) membranes, the numbers of Candida albicans, Pseudomonas aeruginosa and Staphylococcus epidermidis crossing the membranes were an order of magnitude lower when the membranes were punctured by microneedles rather than a 21G hypodermic needle. Apart from the movement of C. albicans across hypodermic needle-punctured membranes, where 40.2% of the microbial load on control membranes permeated the barrier over 24 h, the numbers of permeating microorganisms was less than 5% of the original microbial load on control membranes. Experiments employing excised porcine skin and radiolabelled microorganisms showed that the numbers of microorganisms penetrating skin beyond the stratum corneum were approximately an order of magnitude greater than the numbers crossing SilescolA (R) membranes in the corresponding experiments. Approximately 10(3) cfu of each microorganism adhered to hypodermic needles during insertion. The numbers of microorganisms adhering to MN arrays were an order of magnitude higher in each case. We have shown here that microneedle puncture resulted in significantly less microbial penetration than did hypodermic needle puncture and that no microorganisms crossed the viable epidermis in microneedle-punctured skin, in contrast to needle-punctured skin. Given the antimicrobial properties of skin, it is, therefore, likely that application of microneedle arrays to skin in an appropriate manner would not cause either local or systemic infection in normal circumstances in immune-competent patients. In supporting widespread clinical use of microneedle-based delivery systems, appropriate animal studies are now needed to conclusively demonstrate this in vivo. Safety in patients will be enhanced by aseptic or sterile manufacture and by fabricating microneedles from self-disabling materials (e.g. dissolving or biodegradable polymers) to prevent inappropriate or accidental reuse.
引用
收藏
页码:2513 / 2522
页数:10
相关论文
共 24 条
[1]   Preclinical evaluation of microneedle technology for intradermal delivery of influenza vaccines [J].
Alarcon, Jason B. ;
Hartley, Andrea Waterston ;
Harvey, Noel G. ;
Mikszta, John A. .
CLINICAL AND VACCINE IMMUNOLOGY, 2007, 14 (04) :375-381
[2]   Transdermal delivery of desmopressin using a coated microneedle array patch system [J].
Cormier, M ;
Johnson, B ;
Ameri, M ;
Nyam, K ;
Libiran, L ;
Zhang, DD ;
Daddona, P .
JOURNAL OF CONTROLLED RELEASE, 2004, 97 (03) :503-511
[3]  
Coulman Sion A., 2006, Current Drug Delivery, V3, P65, DOI 10.2174/156720106775197510
[4]  
Del Mar CB, 2001, MED J AUSTRALIA, V175, P342
[5]  
Donnelly RF, 2008, J CONTROL RELEASE, V129, P154, DOI 10.1016/j.jconrel.2008.05.002
[6]   Microneedle Arrays Permit Enhanced Intradermal Delivery of a Preformed Photosensitizer [J].
Donnelly, Ryan F. ;
Morrow, Desmond I. J. ;
McCarron, Paul A. ;
David Woolfson, A. ;
Morrissey, Anthony ;
Juzenas, Petras ;
Juzeniene, Asta ;
Iani, Vladimir ;
McCarthy, Helen O. ;
Moan, Johan .
PHOTOCHEMISTRY AND PHOTOBIOLOGY, 2009, 85 (01) :195-204
[7]  
Elsner Peter, 2006, Curr Probl Dermatol, V33, P35
[8]   MINIATURE PIG AS AN ANIMAL-MODEL TO STUDY PHOTOAGING [J].
FOURTANIER, A ;
BERREBI, C .
PHOTOCHEMISTRY AND PHOTOBIOLOGY, 1989, 50 (06) :771-784
[9]   Clinical administration of microneedles: skin puncture, pain and sensation [J].
Haq, M. I. ;
Smith, E. ;
John, D. N. ;
Kalavala, M. ;
Edwards, C. ;
Anstey, A. ;
Morrissey, A. ;
Birchall, J. C. .
BIOMEDICAL MICRODEVICES, 2009, 11 (01) :35-47
[10]   Microfabricated microneedles: A novel approach to transdermal drug delivery [J].
Henry, S ;
McAllister, DV ;
Allen, MG ;
Prausnitz, MR .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1998, 87 (08) :922-925