'Re-evaluation of appropriate size of the laryngeal mask airway

被引:26
作者
Asai, T
Murao, K
Yukawa, H
Shingu, K
机构
[1] Kansai Med Univ, Dept Anaesthesiol, Moriguchi, Osaka 5708507, Japan
[2] Kansai Med Univ, Dept Otolaryngol, Moriguchi, Osaka 5708507, Japan
关键词
intubation tracheal; equipment; masks anaesthesia;
D O I
10.1093/bja/83.3.478
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have assessed 32 males and 31 females in a randomized, crossover study to see if there was any difference in the correct positioning of the laryngeal mask, optimal ventilation (defined as no gas leak around the mask at an airway pressure of 18 cm H2O) and cuff visibility between sizes 4 and 5 masks in males and sizes 3 and 4 in females. The position of the mask in relation to the glottis was assessed using a fibreoptic bronchoscope. There was no significant difference in correct positioning between the two sizes in either sex. Gas leak was significantly less frequent for a larger than a smaller mask (P<0.01 for both sexes), whereas the cuff was more often seen in the mouth with larger masks (P<0.02 for males and P<0.01 for females). Therefore, larger masks (size 4 in females and size 5 in males) provided a better seal than smaller sizes without worsening the relative position of the mask to the glottis; however, the larger mask came up within the mouth more often, which could interfere with tonsillectomy and could increase the risk of sore throat or lingual nerve damage.
引用
收藏
页码:478 / 479
页数:2
相关论文
共 10 条
[1]   Laryngeal mask airway and lingual nerve injury [J].
Ahmad, NS ;
Yentis, SM .
ANAESTHESIA, 1996, 51 (07) :707-708
[2]   CRICOID PRESSURE IMPEDES PLACEMENT OF THE LARYNGEAL MASK AIRWAY [J].
ASAI, T ;
BARCLAY, K ;
POWER, I ;
VAUGHAN, RS .
BRITISH JOURNAL OF ANAESTHESIA, 1995, 74 (05) :521-525
[3]   Appropriate size and inflation of the laryngeal mask airway [J].
Asai, T ;
Howell, TK ;
Koga, K ;
Morris, S .
BRITISH JOURNAL OF ANAESTHESIA, 1998, 80 (04) :470-474
[4]  
Asai T, 1997, DIFFICULTIES TRACHEA, P197
[5]   An evaluation of the factors influencing selection of the optimal size of laryngeal mask airway in normal adults [J].
Berry, AM ;
Brimacombe, JR ;
McManus, KF ;
Goldblatt, M .
ANAESTHESIA, 1998, 53 (06) :565-570
[6]   Lingual nerve injury associated with laryngeal mask use [J].
Brain, AIJ ;
Howard, D .
ANAESTHESIA, 1998, 53 (07) :713-714
[7]  
Brimacombe JR, 1997, The Laryngeal Mask Airway: A Review and Practical Guide
[8]   SAMPLE-SIZE FOR TESTING DIFFERENCES IN PROPORTIONS FOR THE PAIRED-SAMPLE DESIGN [J].
CONNOR, RJ .
BIOMETRICS, 1987, 43 (01) :207-211
[9]  
Nott MR, 1998, BRIT J ANAESTH, V81, P657, DOI 10.1093/bja/81.4.657
[10]   Selection of the proper size of laryngeal mask airway in adults [J].
Voyagis, GS ;
Batzioulis, PG ;
SechaDoussaitou, PN .
ANESTHESIA AND ANALGESIA, 1996, 83 (03) :663-664