Determination of gradient magnetic field-induced acoustic noise associated with the use of echo planar and three-dimensional, fast spin echo techniques

被引:59
作者
Shellock, FG [1 ]
Ziarati, M [1 ]
Atkinson, D [1 ]
Chen, DY [1 ]
机构
[1] Univ So Calif, Sch Med, Los Angeles, CA USA
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1998年 / 8卷 / 05期
关键词
MR biological effects; MR quality assurance; MR technology;
D O I
10.1002/jmri.1880080522
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to assess gradient magnetic-field-induced acoustic noise levels associated with the use of echo planar imaging (EPI) and three-dimensional fast spin echo (3D-FSE) pulse sequences. Acoustic noise measurements were obtained from two different high field-strength MR systems (1.5 T, Siemens and General Electric Co.) under ambient noise conditions and the use of EPI and 3D-FSE pulse sequences. Parameters were selected to produce "worst case" acoustic noise levels. Acoustic noise recordings were made at the entrance, the center, and at the exit of the magnet bores with a specially designed microphone that was unperturbed by electromagnetic fields. The; highest ambient noise levels (A-weighted scale) were 67 dB (Siemens: the same values were recorded at the center and at the exit) and 78 dB (General Electric Co.: recorded at the exit). The highest acoustic noise levels recorded during activation of the gradient magnetic fields were 114 dB (Siemens) and 115 dB (General Electric Co.) and those occurred at the centers of the MR systems with the use of the EPI technique. Gradient magnetic fields associated with the use of EPI and 3D-FSE techniques produced acoustic noise levels that were within permissible levels recommended by federal guidelines.
引用
收藏
页码:1154 / 1157
页数:4
相关论文
共 13 条
[1]   POTENTIAL HEARING-LOSS RESULTING FROM MR IMAGING [J].
BRUMMETT, RE ;
TALBOT, JM ;
CHARUHAS, P .
RADIOLOGY, 1988, 169 (02) :539-540
[2]  
*DEP HLTH, 1993, GUID MAGN RES DIAGN
[3]  
*FDA, 1988, FED REGISTER, V53, P7575
[4]  
Harris Cyril M., 1979, Handbook of noise control
[5]   ACOUSTIC ANALYSIS OF GRADIENT-COIL NOISE IN MR IMAGING [J].
HURWITZ, R ;
LANE, SR ;
BELL, RA ;
BRANTZAWADZKI, MN .
RADIOLOGY, 1989, 173 (02) :545-548
[6]   ACOUSTIC NOISE-LEVELS DURING MAGNETIC-RESONANCE-IMAGING SCANNING AT 1.5 T [J].
MCJURY, M ;
BLUG, A ;
JOERGER, C ;
CONDON, B ;
WYPER, D .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (796) :413-415
[7]   ACOUSTIC NOISE-LEVELS GENERATED DURING HIGH-FIELD MR-IMAGING [J].
MCJURY, MJ .
CLINICAL RADIOLOGY, 1995, 50 (05) :331-334
[8]  
MELNICK W, 1979, HDB NOISE CONTROL, P1
[9]  
MILLER LE, 1979, HDB NOISE CONTROL, P2
[10]  
ROBINSON DW, 1976, EFFECTS NOISE HEARIN, P383