Determination of heavy metals by electrothermal atomic absorption spectrometry after electrodeposition on a graphite probe

被引:34
作者
Komárek, J [1 ]
Holy, J [1 ]
机构
[1] Masaryk Univ, Dept Analyt Chem, CS-61137 Brno, Czech Republic
关键词
electrothermal atomic absorption spectrometry; electrodeposition; graphite probe; heavy metals;
D O I
10.1016/S0584-8547(99)00037-3
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Traces of heavy metals were separated and preconcentrated electrochemically at a controlled potential on the graphite ridge probe. After the electrolysis, the electrode-probe was inserted in the graphite furnace for atomization of metal deposit by an automatic system. Conditions for the electrodeposition, such as pH of solutions, the deposition potential and concentration of electrolyte, were optimized. Detection limits improved with increased time of electrodeposition and were 16 ng l(-1) Cu, 1.0 ng l(-1) Cd, 6.0 ng l(-1) Pb, 64 ng l(-1) Ni, 14 ng l(-1) Cr(III) and 17 ng l(-1) Cr (VI) for a 10-min deposition. This method was applied for the determination of copper, cadmium, lead, nickel and of chromium species in seawater. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:733 / 738
页数:6
相关论文
共 13 条
[1]   DETERMINATION OF CHROMIUM SPECIATION IN NATURAL-WATERS BY ELECTRODEPOSITION ON GRAPHITE TUBES FOR ELECTROTHERMAL ATOMIZATION [J].
BATLEY, GE ;
MATOUSEK, JP .
ANALYTICAL CHEMISTRY, 1980, 52 (11) :1570-1574
[2]   DETERMINATION OF HEAVY-METALS IN SEAWATER BY ATOMIC-ABSORPTION SPECTROMETRY AFTER ELECTRODEPOSITION ON PYROLYTIC GRAPHITE-COATED TUBES [J].
BATLEY, GE .
ANALYTICAL CHEMISTRY, 1977, 49 (13) :2031-2035
[3]   DETERMINATION OF PLATINUM IN BIOTIC AND ENVIRONMENTAL-SAMPLES BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY AFTER ITS ELECTRODEPOSITION INTO A GRAPHITE TUBE PACKED WITH RETICULATED VITREOUS CARBON [J].
BEINROHR, E ;
LEE, ML ;
TSCHOPEL, P ;
TOLG, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (6-9) :689-692
[4]   ONLINE ELECTROCHEMICAL PRECONCENTRATION OF MANGANESE FOR GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY USING A FLOW-THROUGH ELECTROCHEMICAL-CELL [J].
BEINROHR, E ;
RAPTA, M ;
LEE, ML ;
TSCHOPEL, P ;
TOLG, G .
MIKROCHIMICA ACTA, 1993, 110 (1-3) :1-12
[5]   DETERMINATION OF LEAD BY ELECTROTHERMAL VAPORIZATION MICROWAVE-INDUCED PLASMA-ATOMIC EMISSION-SPECTROMETRY AFTER FLOW-THROUGH ELECTROLYTIC DEPOSITION IN A GRAPHITE TUBE PACKED WITH RETICULATED VITREOUS CARBON [J].
BEINROHR, E ;
BULSKA, E ;
TSCHOPEL, P ;
TOLG, G .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (07) :965-968
[6]  
Bulska E, 1997, ATOM SPECTROSC, V18, P202
[7]   THE APPLICATION OF ELECTRODEPOSITION ON A TUNGSTEN WIRE TO FURNACE ATOMIC-ABSORPTION SPECTROMETRY [J].
CZOBIK, EJ ;
MATOUSEK, JP .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1980, 35 (11-1) :741-751
[8]   PURIFICATION OF REAGENTS AND SEPARATION OF ELEMENT TRACES BY ELECTRODEPOSITION ONTO A GRAPHITE TUBE CATHODE [J].
HOPPSTOCK, K ;
GARTEN, RPH ;
TSCHOPEL, P ;
TOLG, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 343 (9-10) :778-781
[9]   APPLICATION OF ELECTRODEPOSITION TECHNIQUES TO FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY .1. DETERMINATION OF CADMIUM WITH A TUNGSTEN FILAMENT [J].
LUND, W ;
LARSEN, BV .
ANALYTICA CHIMICA ACTA, 1974, 70 (02) :299-310
[10]   Analyte stabilization by electrodeposited palladium modifier for electrothermal atomic absorption spectrometry: Characterization by scanning electron microscopy and anodic stripping voltammetry [J].
Matousek, JP ;
Powell, HKJ .
TALANTA, 1997, 44 (07) :1183-1193