Investigation of the analytical performance of an MDMI-ICP-AES system

被引:30
作者
Lazar, AC [1 ]
Farnsworth, PB [1 ]
机构
[1] BRIGHAM YOUNG UNIV,DEPT CHEM & BIOCHEM,PROVO,UT 84602
关键词
monodisperse dried microparticulate injector (MDMI); analytical performance; washout time; vertical emission profile; plasma cooling;
D O I
10.1366/0003702971941016
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A 40-MHz ICP was equipped with a novel sample introduction system capable of introducing monodisperse droplets into the plasma with 100% efficiency. The analytical performance of the !system was investigated by using two different synchronous detection techniques. The detection limits determined for Ba with ensemble averaging and with boxcar averaging were 0.048 ng/mL and 0.063 ng/mL, respectively. Signal stability in time, effect of the matrix on barium ionic emission intensity, vertical emission profiles from a multicomponent solution, and washout time of the system were also studied.
引用
收藏
页码:617 / 624
页数:8
相关论文
共 21 条
[1]   NOISE-POWER SPECTRA OF OPTICAL AND ACOUSTIC-EMISSION SIGNALS FROM AN INDUCTIVELY COUPLED PLASMA [J].
BELCHAMBER, RM ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (01) :17-27
[2]   EFFECT OF SIGNAL INTEGRATION PERIOD ON MEASUREMENT PRECISION IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY [J].
BELCHAMBER, RM ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (01) :71-74
[3]   USE OF A NOVEL SAMPLE MODULATION TECHNIQUE IN FLAME SPECTROMETRY [J].
BOJOVIC, V ;
ANTICJOV.A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1972, B 27 (09) :385-&
[4]  
BOUMANS PWJ, 1987, INDUCTIVELY COUPLE 1, P166
[5]   A SIMPLE, NON-INVASIVE METHOD FOR THE MEASUREMENT OF GAS-FLOW VELOCITIES IN THE INDUCTIVELY COUPLED PLASMA [J].
CICERONE, MT ;
FARNSWORTH, PB .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (09) :897-907
[6]   VERTICAL AND RADIAL EMISSION PROFILES AND ION-ATOM INTENSITY RATIOS IN INDUCTIVELY COUPLED PLASMAS - THE CONNECTION TO VAPORIZING DROPLETS [J].
FISTER, JC ;
OLESIK, JW .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (6-7) :869-883
[7]   MONODISPERSE DRIED MICROPARTICULATE INJECTOR FOR ANALYTICAL INSTRUMENTATION [J].
FRENCH, JB ;
ETKIN, B ;
JONG, R .
ANALYTICAL CHEMISTRY, 1994, 66 (05) :685-691
[9]  
HIEFTJE GM, 1972, ANAL CHEM, V44, pA69, DOI [10.1021/ac60315a036, 10.1021/ac60315a752]
[10]  
INGLE JD, 1988, SPECTROCHEMICAL ANAL, pCH5