Safer epidural and spinal connectors

被引:20
作者
Lanigan, CJ [1 ]
机构
[1] Univ Hosp Lewisham, Dept Anaesthesia & Pain Management, London SE13 6LH, England
关键词
anaesthetic drug delivery; epidural; medical devices; patient safety; spinal;
D O I
10.1046/j.1365-2044.2002.02601.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Inadvertent cross-connection of epidural local anaesthetic delivery systems with intravenous lines can cause significant morbidity and even death. Inadvertent intrathecal injection of toxic intravenous drugs has already led to many deaths. If epidural and spinal systems had the standard Luer lock connection reversed the problem would be prevented. The practical implications of this idea are explored. Patient safety would be enhanced if this system were adopted.
引用
收藏
页码:567 / 571
页数:5
相关论文
共 18 条
[1]   Medical errors: a common problem - It is time to get serious about them [J].
Alberti, KGMM .
BMJ-BRITISH MEDICAL JOURNAL, 2001, 322 (7285) :501-502
[2]   Delivering safe health care - Safety is a patient's right and the obligation of all health professionals [J].
Barach, P ;
Moss, F .
BRITISH MEDICAL JOURNAL, 2001, 323 (7313) :585-586
[3]   Not again! Preventing errors lies in redesign - not exhortation [J].
Berwick, DM .
BMJ-BRITISH MEDICAL JOURNAL, 2001, 322 (7281) :247-248
[4]   Safety matters [J].
Birks, RJS .
ANAESTHESIA, 2001, 56 (09) :823-824
[5]  
*BRIT STAND I, 1994, 2059411994 BSI EN
[6]  
*BRIT STAND I, 1997, 17071997 BSI EN
[7]   Misconnection misadventure [J].
Hutton, A ;
Christie, I .
ANAESTHESIA, 2001, 56 (10) :1022-1023
[8]  
Lanigan C, 2001, ANAESTHESIA, V56, P586
[9]  
LANIGAN C, 1998, INT J ACUTE PAIN MAN, V1, P7
[10]  
*NAT CEPOD, 2001, CHANG WAY WE OP 2001