The x-ray sensitivity of amorphous selenium for mammography

被引:66
作者
Stone, MF
Zhao, W
Jacak, BV
O'Connor, P
Yu, B
Rehak, P
机构
[1] SUNY Stony Brook, Hlth Sci Ctr 120, Dept Radiol, Stony Brook, NY 11794 USA
[2] SUNY Stony Brook, Dept Phys & Astron, Stony Brook, NY 11794 USA
[3] Brookhaven Natl Lab, Instrumentat Div, Upton, NY 11973 USA
关键词
amorphous selenium; x-ray sensitivity; mammography;
D O I
10.1118/1.1449874
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A study of the x-ray sensitivity of amorphous selenium (a-Se) for digital mammography has been performed. A uniform layer of a-Se was deposited on a glass substrate with electrodes on both surfaces. The deposition procedure was identical to that used for a-Se flat-panel detectors. A high voltage was applied to the top surface of the a-Se layer in order to establish an electric field E-Se. Then the sample was exposed to x rays with 27 kVp spectra generated from an x-ray tube with a molybdenum (Mo) target. The mean x-ray energy of the spectrum used was similar to16.6 keV. The x-ray current generated by the a-Se layer was measured as a function of E-Se. From the current measurement and the estimation of total x-ray energy absorbed in the a-Se, the energy required to create one electron-hole pair (EHP), W, was determined as a function of E-Se. It was found that at the most commonly used E-Se of 10 V/mum, W was measured as 64 eV. This is considerably higher than the widely accepted typical value of W = 50 eV measured at higher x-ray photon energies (e.g., 50 keV). The dependence of W as a function of E-Se can be best fitted using the empirical expression of E-Se(-y). This relationship is consistent with the results obtained at higher x-ray energies. This article provides an accurate measurement of x-ray sensitivity of a-Se at mammographic energies independent of detector operation, such as the most recently developed flat-panel detectors. The results will be a useful toot for investigation and optimization of a-Se-based x-ray imaging detectors, such as determination of pixel fill-factor and optimal E-Se during operation. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:319 / 324
页数:6
相关论文
共 30 条
[1]   X-RAY-SENSITIVITY OF HALOGENATED A-SE-AS PHOTORECEPTORS FOR ELECTRORADIOGRAPHY [J].
AIYAH, V ;
BAILLIE, A ;
POLISCHUK, B ;
BEKIROV, A ;
KASAP, SO .
JOURNAL OF NON-CRYSTALLINE SOLIDS, 1991, 137 :1337-1340
[2]   An integrated CMOS-selenium x-ray detector for digital mammography [J].
Andre, MP ;
Spivey, B ;
Martin, P ;
Morsell, L ;
Atlas, E ;
Pellegrino, T .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :204-209
[3]   DEMONSTRATION OF MEGAVOLTAGE AND DIAGNOSTIC-X-RAY IMAGING WITH HYDROGENATED AMORPHOUS-SILICON ARRAYS [J].
ANTONUK, LE ;
BOUDRY, J ;
HUANG, WD ;
MCSHAN, DL ;
MORTON, EJ ;
YORKSTON, J ;
LONGO, MJ ;
STREET, RA .
MEDICAL PHYSICS, 1992, 19 (06) :1455-1466
[4]  
BEVINGTON R, 1992, DATA REDUCTION ERROR, P55
[5]   Measurement of x-ray photogeneration in amorphous selenium [J].
Blevis, IM ;
Hunt, DC ;
Rowlands, JA .
JOURNAL OF APPLIED PHYSICS, 1999, 85 (11) :7958-7963
[6]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[7]   Operation of amorphous silicon detectors for chest radiography within system latency requirements [J].
Boudry, JM .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :336-344
[8]   ENERGY AND ANGULAR-DEPENDENCE OF X-RAY ABSORPTION AND ITS EFFECT ON RADIOGRAPHIC RESPONSE IN SCREEN-FILM SYSTEMS [J].
CHAN, HP ;
DOI, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (05) :565-579
[9]   New CsI/a-Si 17" x 17" X-ray flat panel detector provides superior detectivity and immediate direct digital output for General Radiography systems [J].
Chaussat, C ;
Chabbal, J ;
Ducourant, T ;
Spinnler, V ;
Vieux, G ;
Neyret, R .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :45-56
[10]  
Cheung L, 1998, COMP IMAG VIS, V13, P11