Evaluation of the revised MicroScan Dried Overnight Gram-Positive Identification panel to identify Enterococcus species
被引:8
作者:
Iwen, PC
论文数: 0引用数: 0
h-index: 0
机构:
Univ Nebraska, Med Ctr, Dept Pathol & Microbiol, Nebraska Med Ctr 986495, Omaha, NE 68198 USAUniv Nebraska, Med Ctr, Dept Pathol & Microbiol, Nebraska Med Ctr 986495, Omaha, NE 68198 USA
Iwen, PC
[1
]
Rupp, ME
论文数: 0引用数: 0
h-index: 0
机构:Univ Nebraska, Med Ctr, Dept Pathol & Microbiol, Nebraska Med Ctr 986495, Omaha, NE 68198 USA
Rupp, ME
Schreckenberger, PC
论文数: 0引用数: 0
h-index: 0
机构:Univ Nebraska, Med Ctr, Dept Pathol & Microbiol, Nebraska Med Ctr 986495, Omaha, NE 68198 USA
Schreckenberger, PC
Hinrichs, SH
论文数: 0引用数: 0
h-index: 0
机构:Univ Nebraska, Med Ctr, Dept Pathol & Microbiol, Nebraska Med Ctr 986495, Omaha, NE 68198 USA
Hinrichs, SH
机构:
[1] Univ Nebraska, Med Ctr, Dept Pathol & Microbiol, Nebraska Med Ctr 986495, Omaha, NE 68198 USA
[2] Univ Nebraska, Med Ctr, Dept Healthcare Epidemiol, Omaha, NE USA
[3] Univ Illinois, Coll Med, Dept Pathol, Chicago, IL 60612 USA
The revised MicroScan Dried Overnight Gram-Positive Identification panel was evaluated for its efficacy at identifying Enterococcus species in comparison with conventional biochemical tests. Supplemental testing of ampicillin-susceptible Enterococcus faecium for motility and the ability to acidify methyl-alpha-D-glucopyranoside helped recognize E. gallinarum and increased the accuracy of the panel for identifying Enterococcus species to 98.5%.