A Simple Sign for Recognizing Off-Axis OCT Measurement Beam Placement in the Context of Multicentre Studies

被引:16
作者
Balk, Lisanne J. [1 ]
de Vries-Knoppert, Willemien A. E. J. [2 ]
Petzold, Axel [1 ,3 ]
机构
[1] Vrije Univ Amsterdam Med Ctr, MS Ctr Amsterdam, Amsterdam, Netherlands
[2] Vrije Univ Amsterdam Med Ctr, Dept Ophthalmol, Amsterdam, Netherlands
[3] UCL Inst Neurol, London, England
关键词
OPTICAL COHERENCE TOMOGRAPHY; FIBER LAYER THICKNESS; MULTIPLE-SCLEROSIS; QUALITY; VISION;
D O I
10.1371/journal.pone.0048222
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
070301 [无机化学]; 070403 [天体物理学]; 070507 [自然资源与国土空间规划学]; 090105 [作物生产系统与生态工程];
摘要
Purpose: Optical coherence tomography (OCT) allows quantification of the thickness of the retinal nerve fibre layer (RNFL) thickness, a potential biomarker for neurodegeneration. The estimated annual RNFL loss in multiple sclerosis amounts to 2 mm using time domain OCT. The recognition of measurement artifacts exceeding this limit is relevant for the successful use of OCT as a secondary outcome measure in clinical trials. Methods: Prospective study design. An exploratory pilot study (ring and volume scans) followed by a cohort study (1,980 OCT ring scans). The OCT measurement beam was placed off-axis to the left, right, top and bottom of the subjects pupil and RNFL thickness of these scans were compared to the centrally placed reference scans. Results: Off-axis placement of the OCT measurement beam resulted in significant artifacts in RNFL thickness measurements (95%CI 9 mu m, maximal size of error 42 mu m). Off-axis placement gave characteristic patterns of the OCT live images which are not necessarily saved for review. Off-axis placement also causes regional inhomogeneity of reflectivity in the outer nuclear (ONL) and outer plexiform layers (OPL) which remains visible on scans saved for review. Conclusion: Off-axis beam placement introduces measurement artifacts at a magnitude which may mask recognition of RNFL loss due to neurodegeneration in multiple sclerosis. The resulting pattern in the OCT live image can only be recognised by the technician capturing the scans. Once the averaged scans have been aligned this pattern is lost. Retrospective identification of this artifact is however possible by presence of regional inhomogeneity of ONL/OPL reflectivity. This simple and robust sign may be considered for quality control criteria in the setting of multicentre OCT studies. The practical advice of this study is to keep the OCT image in the acquisition window horizontally aligned whenever possible.
引用
收藏
页数:8
相关论文
共 10 条
[1]
Effect of image quality on tissue thickness measurements obtained with spectral domain-optical coherence tomography [J].
Balasubramanian, Madhusudhanan ;
Bowd, Christopher ;
Vizzeri, Gianmarco ;
Weinreb, Robert N. ;
Zangwill, Linda M. .
OPTICS EXPRESS, 2009, 17 (05) :4019-4036
[2]
Imaging outcomes for neuroprotection and repair in multiple sclerosis trials [J].
Barkhof, Frederik ;
Calabresi, Peter A. ;
Miller, David H. ;
Reingold, Stephen C. .
NATURE REVIEWS NEUROLOGY, 2009, 5 (05) :256-266
[3]
QUALITY ISSUES IN INTERPRETATION OF OPTICAL COHERENCE TOMOGRAMS IN MACULAR DISEASES [J].
Domalpally, Amitha ;
Danis, Ronald P. ;
Zhang, Baoyan ;
Myers, Dawn ;
Kruse, Christina N. .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2009, 29 (06) :775-781
[4]
Revealing Henle's Fiber Layer Using Spectral Domain Optical Coherence Tomography [J].
Lujan, Brandon J. ;
Roorda, Austin ;
Knighton, Robert W. ;
Carroll, Joseph .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (03) :1486-1492
[5]
IMPROVED VISUALIZATION OF HENLE FIBER LAYER BY CHANGING THE MEASUREMENT BEAM ANGLE ON OPTICAL COHERENCE TOMOGRAPHY [J].
Otani, Tomohiro ;
Yamaguchi, Yumiko ;
Kishi, Shoji .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2011, 31 (03) :497-501
[6]
THE LENGTHS OF THE FIBERS OF HENLE IN THE RETINA OF MACAQUE MONKEYS - IMPLICATIONS FOR VISION [J].
PERRY, VH ;
COWEY, A .
NEUROSCIENCE, 1988, 25 (01) :225-236
[7]
Petzold A, 2010, LANCET NEUROL, V9, P921, DOI 10.1016/S1474-4422(10)70168-X
[8]
THE ANATOMICAL BASIS OF THE CAECOCENTRAL SCOTOMA - NEW OBSERVATIONS AND A REVIEW [J].
PLANT, GT ;
PERRY, VH .
BRAIN, 1990, 113 :1441-1457
[9]
Longitudinal Study of Vision and Retinal Nerve Fiber Layer Thickness in Multiple Sclerosis [J].
Talman, Lauren S. ;
Bisker, Esther R. ;
Sackel, David J. ;
Long, David A., Jr. ;
Galetta, Kristin M. ;
Ratchford, John N. ;
Lile, Deacon J. ;
Farrell, Sheena K. ;
Loguidice, Michael J. ;
Remington, Gina ;
Conger, Amy ;
Frohman, Teresa C. ;
Jacobs, Dina A. ;
Markowitz, Clyde E. ;
Cutter, Gary R. ;
Ying, Gui-Shuang ;
Dai, Yang ;
Maguire, Maureen G. ;
Galetta, Steven L. ;
Frohman, Elliot M. ;
Calabresi, Peter A. ;
Balcer, Laura J. .
ANNALS OF NEUROLOGY, 2010, 67 (06) :749-760
[10]
Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography [J].
Wu, Huijuan ;
de Boer, Johannes F. ;
Chen, Teresa C. .
JOURNAL OF GLAUCOMA, 2011, 20 (08) :470-476