Co-registration of magnetoencephalography with magnetic resonance imaging using bite-bar-based fiducials and surface-matching

被引:80
作者
Adjamian, P [1 ]
Barnes, GR
Hillebrand, A
Holliday, IE
Singh, KD
Furlong, PL
Harrington, E
Barclay, CW
Route, PJG
机构
[1] Aston Univ, Wellcome Trust Lab MEG Studies, Neurosci Res Inst, Birmingham B4 7ET, W Midlands, England
[2] Univ Birmingham, Sch Dent, Birmingham, W Midlands, England
关键词
MEG-MRI co-registration; bite-bar; fiducial localization error; target registration error; Monte Carlo simulation;
D O I
10.1016/j.clinph.2003.10.023
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To introduce a new technique for co-registration of Magnetoencephalography (MEG) with magnetic resonance imaging (MRI). We compare the accuracy of a new bite-bar with fixed fiducials to a previous technique whereby fiducial coils were attached proximal to landmarks on the skull. Methods: A bite-bar with fixed fiducial coils is used to determine the position of the head in the MEG co-ordinate system. Co-registration is performed by a surface-matching technique. The advantage of fixing the coils is that the co-ordinate system is not based upon arbitrary and operator dependent fiducial points that are attached to landmarks (e.g. nasion and the preauricular points), but rather on those that are permanently fixed in relation to the skull. Results: As a consequence of minimizing coil movement during digitization, errors in localization of the coils are significantly reduced, as shown by a randomization test. Displacement of the bite-bar caused by removal and repositioning between MEG recordings is minimal (similar to0.5 mm), and dipole localization accuracy of a somatosensory mapping paradigm shows a repeatability of similar to 5 mm. The overall accuracy of the new procedure is greatly improved compared to the previous technique. Conclusions: The test-retest reliability and accuracy of target localization with the new design is superior to techniques that incorporate anatomical-based fiducial points or coils placed on the circumference of the head. (C) 2003 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:691 / 698
页数:8
相关论文
共 13 条
  • [1] AHLFORS S, 1990, ADV BIOMAGNETISM
  • [3] The use of an MEG device as 3D digitizer and motion monitoring system
    de Munck, JC
    Verbunt, JPA
    Van't Ent, D
    Van Dijk, BW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (08) : 2041 - 2052
  • [4] Predicting error in rigid-body point-based registration
    Fitzpatrick, JM
    West, JB
    Maurer, CR
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 694 - 702
  • [5] GEORGE JS, 1990, ADV BIOMAGNETISM
  • [6] HILLEBRAND A, 2000, THESIS ASTON U BIRMI
  • [7] Estimation of the accuracy of a surface matching technique for registration of EEG and MRI data
    Huppertz, HJ
    Otte, M
    Grimm, C
    Kristeva-Feige, R
    Mergner, T
    Lücking, CH
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1998, 106 (05): : 409 - 415
  • [8] Multidimensional alignment using the Euclidean distance transform
    Kozinska, D
    Tretiak, OJ
    Nissanov, J
    Ozturk, C
    [J]. GRAPHICAL MODELS AND IMAGE PROCESSING, 1997, 59 (06): : 373 - 387
  • [9] Registration of head CT images to physical space using a weighted combination of points and surfaces
    Maurer, CR
    Maciunas, RJ
    Fitzpatrick, JM
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 753 - 761
  • [10] ACCURATE 3-DIMENSIONAL REGISTRATION OF CT, PET, AND OR MR IMAGES OF THE BRAIN
    PELIZZARI, CA
    CHEN, GTY
    SPELBRING, DR
    WEICHSELBAUM, RR
    CHEN, CT
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1989, 13 (01) : 20 - 26