Thermal softening of tracheal tubes: an unrecognized hazard of the Bair Hugger active patient warming system

被引:18
作者
Ayala, JL
Coe, A
机构
[1] Department of Anaesthesia, Scunthorpe and Goole Hospitals, NHS Trust, Scunthorpe DN15 7BH, Cliff Gardens
关键词
equipment; warming devices; tubes tracheal;
D O I
10.1093/bja/79.4.543
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The Bair Hugger system is a new and highly effective active patient warming system which produces a layer of warm air between the patient and the warming system. We report an instance of marked softening and distortion of a polyvinyl chloride tracheal tube caused by this layer. We also present laboratory data indicating that this is a likely problem under routine theatre conditions, with suggestions for prevention.
引用
收藏
页码:543 / 545