Determination of atrazine using square wave voltammetry with the Hanging Mercury Drop Electrode (HMDE)

被引:71
作者
dos Santos, LBO [1 ]
Abate, G [1 ]
Masini, JC [1 ]
机构
[1] Univ Sao Paulo, Inst Quim, BR-05513970 Sao Paulo, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
square wave voltammetry; atrazine; mercury electrode; water;
D O I
10.1016/j.talanta.2003.08.034
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper presents the optimization of instrumental and solution parameters for determination of atrazine in river waters and formulation by square wave voltammetry (SWV) using a hanging mercury drop electrode. The best sensitivity (35.2+/-0.4 muA ml mug(-1)) was achieved using a frequency of 400 Hz and a medium composed of 40 mmol l(-1) Britton-Robinson (BR) buffer at pH 1.9. The detection limit was 2 mug l(-1) with a linear dynamic range between 10 and 250 mug l(-1). Application of the method to real samples of river waters fortified with 10 mug l(-1) I of atrazine resulted recoveries between 92 and 116%. Additionally, good agreement was observed between results obtained by the proposed method and by HPLC for river water samples spiked with 25 mug l(-1) of atrazine. The determination was not affected by the presence of humic acid at concentration of 5 mg l(-1), indicating that interactions of the herbicide with this class of compounds are fully labile. The stability of the voltammetric signal for samples spiked with 250 mug l(-1) atrazine was evaluated over a period of 14 days in four samples. For two samples, no systematic variation was observed, while for the other two, a decrease of peak current between 3 and 15% occurred, suggesting that the stability is dependent on the sample nature. HPLC analyses suggest formation of deethylatrazine during the second week of storage in the samples for which the SWV peak current had the more intense decrease. (C). 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:667 / 674
页数:8
相关论文
共 31 条
[1]   Stripping voltammetry for the determination of trace metal speciation and in-situ measurements of trace metal distributions in marine waters [J].
Achterberg, EP ;
Braungardt, C .
ANALYTICA CHIMICA ACTA, 1999, 400 :381-397
[2]   DETERMINATION OF ATRAZINE AND SIMAZINE IN ESTUARINE SAMPLES BY HIGH-RESOLUTION GAS-CHROMATOGRAPHY AND NITROGEN SELECTIVE DETECTION [J].
AHEL, M ;
EVANS, KM ;
FILEMAN, TW ;
MANTOURA, RFC .
ANALYTICA CHIMICA ACTA, 1992, 268 (02) :195-204
[3]  
ARMSTRONG DE, PESTICIDES SOIL WATE, P123
[4]   ENVIRONMENTAL-PROTECTION-AGENCY AND OTHER METHODS FOR THE DETERMINATION OF PRIORITY PESTICIDES AND THEIR TRANSFORMATION PRODUCTS IN WATER [J].
BARCELO, D .
JOURNAL OF CHROMATOGRAPHY, 1993, 643 (1-2) :117-143
[5]   SQUARE-WAVE POLAROGRAPHY [J].
BARKER, GC ;
JENKINS, IL .
ANALYST, 1952, 77 (920) :685-696
[6]   Voltammetric behavior of sildenafil citrate (Viagra) using square wave and adsorptive stripping square wave techniques -: Determination in pharmaceutical products [J].
Berzas, JJ ;
Rodriguez, J ;
Castañeda, G ;
Villaseñor, MJ .
ANALYTICA CHIMICA ACTA, 2000, 417 (02) :143-148
[7]  
BORMAN S, 1982, ANAL CHEM, V54, pA698
[8]  
dosSantos MMC, 1996, J ELECTROANAL CHEM, V413, P97
[9]   Cathodic stripping square wave voltammetry of Cu(II)oxine complexes. A mechanistic study [J].
Garay, F ;
Solis, V ;
Lovric, M .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1999, 478 (1-2) :17-24
[10]   Impacts of atrazine in aquatic ecosystems [J].
Graymore, M ;
Stagnitti, F ;
Allinson, G .
ENVIRONMENT INTERNATIONAL, 2001, 26 (7-8) :483-495