Diagnostic performance of a dual-chamber cardioverter defibrillator programmed with nominal settings: A European prospective study

被引:23
作者
Sadoul, N
Jung, W
Jordaens, L
Leenhardt, A
Santini, M
Wolpert, C
Aliot, E
机构
[1] CHU Nancy Brabois, Serv Cardiol, F-54511 Vandoeuvre Les Nancy, France
[2] Univ Bonn Klin, Bonn, Germany
[3] Univ Ziekenhuis, Ghent, Belgium
[4] Hop Univ Lariboisiere, Paris, France
[5] Osped S Filippo Neri, Rome, Italy
关键词
implantable cardioverter defibrillator; dual chamber; detection algorithm;
D O I
10.1046/j.1540-8167.2002.00025.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Despite technologic developments, accurate discrimination of ventricular tachyarrhythmia from rapid rhythms of nonventricular origin remains a challenge. We sought to examine the sensitivity and specificity of a dual-chamber arrhythmia detection algorithm, the PARAD algorithm, incorporated in a dual-chamber implantable cardioverter defibrillator, the Defender (ELA Medical). Methods and Results : All detailed tachycardia episodes (i.e., with stored atrial and ventricular channel markers and electrograms) retrieved from the Holter memory of the device were analyzed from 95 patients (86 men and 9 women; age 62 +/- 12 years) implanted with the Defender with the algorithm programmed at nominal settings. Over a follow-up of 15 +/- 8 months, 559 detailed sustained tachycardia episodes detected in the tachycardia zone were gathered in 62 patients. Of the 300 ventricular tachycardia (VT) episodes, 298 were appropriately classified as VT by the algorithm. Of the 259 supraventricular tachycardia (SVT) episodes, 231 were appropriately classified as SVT. In 8 patients, 28 episodes of SVT were misclassified as VT: 25 atrial fibrillation episodes (6 patients), 2 sinus tachycardia (1 patient), and 1 atrial tachycardia. Calculated sensitivity and specificity were 99.3% and 89.2% on a per episode basis and 99.8% (95% confidence interval: 97.8 to 100) and 91.6% (95% confidence interval: 86.0 to 97.3) on a per patient basis, respectively. Conclusion: In this selected population of patients, the PARAD algorithm was safe and reliable for detection of a wide spectrum of tachyarrhythmias. Its specificity was high, particularly with respect to sinus tachycardia detection, but it must be improved with respect to detection of atrial fibrillation.
引用
收藏
页码:25 / 32
页数:8
相关论文
共 19 条
[1]   Enhanced detection criteria in implantable defibrillators [J].
Brugada, J ;
Mont, L ;
Figueiredo, M ;
Valentino, M ;
Matas, M ;
Navarro-López, F .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (03) :261-268
[2]   Augmented two-channel arrhythmia detection: An efficient diagnostic method for implantable devices [J].
Chiang, CMJ ;
Jenkins, JM ;
Caswell, SA ;
Stevenson, SA ;
DiCarlo, LA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1493-1501
[3]   Variability of atrial sensing during exercise in a patient with a dual chamber ICD caused inappropriate ICD discharges [J].
Glotzer, TV ;
Radoslovich, GA ;
Neglia, J ;
Zimmerman, JM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (12) :1838-1841
[4]   ELECTROCARDIOGRAPHICALLY DOCUMENTED UNNECESSARY, SPONTANEOUS SHOCKS IN 241 PATIENTS WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATORS [J].
GRIMM, W ;
FLORES, BF ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1667-1673
[5]   Stability: An ICD detection criterion for discriminating atrial fibrillation from ventricular tachycardia [J].
Higgins, SL ;
Lee, RS ;
Kramer, RL .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1995, 6 (12) :1081-1088
[6]   Clinical experience with the new detection algorithms for atrial fibrillation of a defibrillator with dual chamber sensing and pacing [J].
Kühlkamp, V ;
Dörnberger, V ;
Mewis, C ;
Suchalla, R ;
Bosch, RF ;
Seipel, L .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1999, 10 (07) :905-915
[7]   Preliminary clinical experience with the first dual chamber pacemaker defibrillator [J].
Lavergne, T ;
Daubert, JC ;
Chauvin, M ;
Dolla, E ;
Kacet, S ;
Leenhardt, A ;
Mabo, P ;
Ritter, P ;
Sadoul, N ;
Saoudi, N ;
Henry, C ;
Nitzsche, R ;
Ripart, A ;
Murgatroyd, F .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01) :182-188
[8]  
Le Franc P, 1997, PACE, V20, P2882
[9]   VENTRICULOATRIAL CONDUCTION IN PATIENTS WITH IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS - IMPLICATIONS FOR TACHYCARDIA DISCRIMINATION BY DUAL-CHAMBER SENSING [J].
LI, HG ;
THAKUR, RK ;
YEE, R ;
KLEIN, GJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (12) :2304-2306
[10]  
Nair M, 1997, CIRCULATION, V95, P967