Correction of physiologically induced global off resonance effects in dynamic echo-planar and spiral functional imaging

被引:168
作者
Pfeuffer, J
Van de Moortele, PF
Ugurbil, K
Hu, XP
Glover, GH
机构
[1] Univ Minnesota, Sch Med, Dept Radiol, Ctr Magnet Resonance Res, Minneapolis, MN 55455 USA
[2] Stanford Univ, Sch Med, Dept Radiol, Stanford, CA USA
关键词
respiration artifacts; physiological artifacts; global off-resonance; functional magnetic resonance imaging;
D O I
10.1002/mrm.10065
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In functional magnetic resonance imaging, a rapid method such as echo-planar (EPI) or spiral is used to collect a dynamic series of images. These techniques are sensitive to changes in resonance frequency which can arise from respiration and are more significant at high magnetic fields. To decrease the noise from respiration-induced phase and frequency fluctuations, a simple correction of the "dynamic off-resonance in k-space" (DORK) was developed. The correction uses phase information from the center of k-space and a navigator echo and is illustrated with dynamic scans of single-shot and segmented EPI and, for the first time, spiral imaging of the human brain at 7 T. Image noise in the respiratory spectrum was measured with an edge operator. The DORK correction significantly reduced respiration-induced noise (image shift for EPI, blurring for spiral, ghosting for segmented acquisition). While spiral imaging was found to exhibit less noise than EPI before correction, the residual noise after the DORK correction was comparable. The correction is simple to apply and can correct for other sources of frequency drift and fluctuations in dynamic imaging. Magn Reson Med 47:344-353, 2002. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:344 / 353
页数:10
相关论文
共 35 条
  • [1] ADRIANY G, 2001, P 9 SCI M ISMRM GLAS, P1097
  • [2] MULTI-PLANAR IMAGE-FORMATION USING NMR SPIN ECHOES
    MANSFIELD, P
    [J]. JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1977, 10 (03): : L55 - L58
  • [3] IMAGE-RECONSTRUCTION FOR ECHO PLANAR IMAGING WITH NONEQUIDISTANT KAPPA-SPACE SAMPLING
    BRUDER, H
    FISCHER, H
    REINFELDER, HE
    SCHMITT, F
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (02) : 311 - 323
  • [4] IMPACT: Image-based physiological artifacts estimation and correction technique for functional MRI
    Chuang, KH
    Chen, JH
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (02) : 344 - 353
  • [5] Artifact due to Bo fluctuations in fMRI:: Correction using the k-space central line
    Durand, E
    van de Moortele, PF
    Pachot-Clouard, M
    Le Bihan, D
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (01) : 198 - 201
  • [6] Estimation of respiration-induced noise fluctuations from undersampled multislice fMRI data
    Frank, LR
    Buxton, RB
    Wong, EC
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (04) : 635 - 644
  • [7] Self-navigated spiral fMRI: Interleaved versus single-shot
    Glover, GH
    Lai, S
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (03) : 361 - 368
  • [8] Glover GH, 1999, MAGNET RESON MED, V42, P412, DOI 10.1002/(SICI)1522-2594(199908)42:2<412::AID-MRM25>3.0.CO
  • [9] 2-U
  • [10] Glover GH, 2000, MAGNET RESON MED, V44, P162, DOI 10.1002/1522-2594(200007)44:1<162::AID-MRM23>3.0.CO