Imaging technique implemented in CellTracks system

被引:14
作者
Tibbe, AGJ
de Grooth, BG
Greve, J
Dolan, GJ
Terstappen, LWMM
机构
[1] Univ Twente, Fac Appl Phys, Biophys Tech Grp, NL-7500 AE Enschede, Netherlands
[2] Immunicon Corp, Huntingdon Valley, PA USA
来源
CYTOMETRY | 2002年 / 47卷 / 04期
关键词
image analysis; high resolution; cell analyzer; CellTracks system;
D O I
10.1002/cyto.10085
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: We developed the CellTracks cell analysis system that, similar to flow cytometry, yields multiparameter information by which the cells can be differentiated. We describe the implementation of a laser scanning imaging method in the system. Image analysis of the cells improves the specificity of cell classification, especially in cases where the particular cells are found relatively infrequently and one has to discriminate between artifacts and real events. Methods: Fluorescent images of immunomagnetically labeled and aligned cells are obtained by passing the bells through a laser focus. The laser focus is smaller than the objects and subsequent frames captured by a regular surveillance CCD camera with a frame grabber board represent different parts of the cells. Complete images of the cells are constructed by shifting each image with respect to each other and adding individual pixel values. Results: The power of combining a fluorescent image with multiparametric data is demonstrated by imaging fluorescent and magnetically labeled beads and cells. The image gives additional information about the dye distribution across the objects. Changes in dye distribution as a function of time were observed in leukocytes labeled with the red fluorescent label, Oxazine750, which are imaged at different time intervals. Conclusions: An imaging technique implemented in the CellTracks system provides high-resolution fluorescent images of events previously identified by the system. The images of the fluorescent cells enhance the ability to classify rare events. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:248 / 255
页数:8
相关论文
共 5 条
[1]   FLOW-CYTOMETRY OF DNA CONTENT USING OXAZINE-750 OR RELATED LASER-DYES WITH 633-NM EXCITATION [J].
SHAPIRO, HM ;
STEPHENS, S .
CYTOMETRY, 1986, 7 (01) :107-110
[2]   Optical tracking and detection of immunomagnetically selected and aligned cells [J].
Tibbe, AGJ ;
de Grooth, BG ;
Greve, J ;
Liberti, PA ;
Dolan, GJ ;
Terstappen, LWMM .
NATURE BIOTECHNOLOGY, 1999, 17 (12) :1210-1213
[3]  
Tibbe AGJ, 2001, CYTOMETRY, V43, P31, DOI 10.1002/1097-0320(20010101)43:1<31::AID-CYTO1016>3.0.CO
[4]  
2-F
[5]  
TIBBE AGJ, IN PRESS CYTOMETRY