Comparison of Time-Domain Short-Term Heart Interval Variability Analysis Using a Wrist-Worn Heart Rate Monitor and the Conventional Electrocardiogram

被引:87
作者
Grossi Porto, Luiz Guilherme [1 ]
Junqueira, Luiz F., Jr.
机构
[1] Univ Brasilia, Fac Med, Cardiovasc Lab, Clin Med Area, BR-70910900 Brasilia, DF, Brazil
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2009年 / 32卷 / 01期
关键词
wrist-worn heart rate monitor; Polar monitor; exercise practice; heart rate variability; cardiac autonomic function; SYMPATHOVAGAL BALANCE; PHYSIOLOGICAL-BASIS; NEURAL MECHANISMS; EXERCISE; ACCURACY; VALIDITY; DISEASE; MODULATION; AGREEMENT; RISK;
D O I
10.1111/j.1540-8159.2009.02175.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Wrist-worn heart rate monitors have not been extensively validated for heart rate variability analysis. The purpose of this study was to compare time-domain variability of heart interval series (R-Ri) recorded by the Polar S810 monitor (Polar Electro Oy, Kempele, Finland) and the conventional electrocardiogram (ECG). Methods: Agreement was verified between variability indices of 5-minute R-Ri simultaneously recorded by both devices and processed by unique software, from 33 subjects aged 18 to 42 years, normal or with different clinical conditions, in rest supine and active standing. ECG minus Polar differences were quantified by the Bland-Altman analysis, and tested by the one-sample t-test or Wilcoxon test. Results: In the supine position, the Polar overestimates (P < 0.0001) the absolute and percentage mean or median of the number (-2.00; -0.49%) and mean of R-Ri (-1.85 ms; -0.20%) and pNN50 (-2.20%; -8.68%), and underestimates the standard deviation (SDNN) (0.32 ms; 0.59%) (P = 0.08; P = 0.02) and root mean square successive difference (RMSSD) (0.90 ms; 1.56%) (P = 0.0008; P < 0.0001). The coefficient of variation (CV) showed null difference. On standing, differences were overestimated for the number (-2.61 intervals; -0.64%) and mean of R-Ri (-0.70 ms; -0.09%), and underestimated for rMSSD (1.70 ms; 10.84%) (P < 0.0001 to < 0.02). The SDNN, CV, and pNN50 indices did not show differences (P = 0.12 to 0.73). Conclusions: The Polar S810 monitor was feasible and reliable for recording short-term R-R interval series, showing excellent agreement with the ECG in providing the time-domain indexes of heart interval variability with differences functionally not relevant. The CV showed the higher agreement in both postures, and the SDNN and pNN50 in the standing posture. (PACE 2009; 32:43-51).
引用
收藏
页码:43 / 51
页数:9
相关论文
共 51 条
[1]   BEAT TO BEAT VARIABILITY IN CARDIOVASCULAR VARIABLES - NOISE OR MUSIC [J].
APPEL, ML ;
BERGER, RD ;
SAUL, JP ;
SMITH, JM ;
COHEN, RJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (05) :1139-1148
[2]   Heart rate variability in athletes [J].
Aubert, AE ;
Seps, B ;
Beckers, F .
SPORTS MEDICINE, 2003, 33 (12) :889-919
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Heart rate monitors and abnormal heart rhythm detection [J].
Boudet, G ;
Chamoux, A .
ARCHIVES OF PHYSIOLOGY AND BIOCHEMISTRY, 2000, 108 (04) :371-379
[5]  
Boudet G, 2001, J SPORT MED PHYS FIT, V41, P546
[6]   ASSOCIATION BETWEEN HEART-RATE-VARIABILITY AND TRAINING RESPONSE IN SEDENTARY MIDDLE-AGED MEN [J].
BOUTCHER, SH ;
STEIN, P .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1995, 70 (01) :75-80
[7]  
Camm AJ, 1996, CIRCULATION, V93, P1043
[8]  
CARROLL T, 1991, Medicine and Science in Sports and Exercise, V23, pS14
[9]   Cardiac autonomic modulation evaluated by heart interval variability is unaltered but subtly widespread in the indeterminate Chagas' disease [J].
Correia, Dalmo ;
Junqueira, Luiz Fernando, Jr. ;
Molina, Rodrigo Juliano ;
Prata, Aluizio .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2007, 30 (06) :772-780
[10]  
de Carvalho JLA, 2002, 2002 6TH INTERNATIONAL CONFERENCE ON SIGNAL PROCESSING PROCEEDINGS, VOLS I AND II, P1488, DOI 10.1109/ICOSP.2002.1180076