Enhanced accuracy of defect detection by myocardial single-photon emission computed tomography with attenuation correction with gadolinium 153 line sources: Evaluation with a cardiac phantom

被引:14
作者
He, ZX
Scarlett, MD
Mahmarian, JJ
Verani, MS
机构
[1] BAYLOR COLL MED,CARDIOL SECT,HOUSTON,TX 77030
[2] METHODIST HOSP,HOUSTON,TX 77030
关键词
single-photon emission computed tomography; attenuation correction; artifacts;
D O I
10.1016/S1071-3581(97)90080-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Photon attenuation is a major cause of artifacts on single-photon emission tomographic imaging, Methods and Results, To study a new method to perform photon attenuation correction (AC), we used a cardiac phantom filed with Tc-99m and imaged it (1) without extrinsic attenuation or defects, (2) with extrinsic attenuation but without defects, (3) without extrinsic attenuation but with defects involving 10% of the myocardial volume and activity ranging from 0% to 75% of maximum, and (4) with both extrinsic attenuation and defects, Transmission and emission images acquired with a dual-head single-photon emission computed tomographic system with Gd-153 line sources were processed by iterative maximum-likelihood-expectation-maximization and were evaluated both qualitatively and quantitatively. The small defects were readily identified both before and after AC. Mean count activity (percent of maximal activity) of the segments with overlying extrinsic attenuation but without defects was only 56% +/- 4% without AC but increased to 86% +/- 4% with AC (p < 0.0001). Without AC, the count activities in the defects with overlying extrinsic attenuation were lower than the actual defect activities, but AC resulted in better approximation of actual defect activities in all but the most severe (0% tracer activity) defects. Conclusion. This new AC method provided an improved estimation of actual myocardial count activity. Even small defects with mild reduction in tracer activity were still identifiable after AC.
引用
收藏
页码:202 / 210
页数:9
相关论文
共 42 条
[1]  
BAILEY DL, 1987, J NUCL MED, V28, P844
[2]  
BELLER GA, 1991, CIRCULATION, V84, P11
[3]  
BONING G, 1995, J NUCL MED S, V36
[5]   ASYMMETRIC FAN TRANSMISSION CT ON SPECT SYSTEMS [J].
CHANG, W ;
LONCARIC, S ;
HUANG, G ;
SANPITAK, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) :913-928
[6]   QUANTITATIVE ROTATIONAL TL-201 TOMOGRAPHY FOR IDENTIFYING AND LOCALIZING CORONARY-ARTERY DISEASE [J].
DEPASQUALE, EE ;
NODY, AC ;
DEPUEY, EG ;
GARCIA, EV ;
PILCHER, G ;
BREDLAU, C ;
ROUBIN, G ;
GOBER, A ;
GRUENTZIG, A ;
DAMATO, P ;
BERGER, HJ .
CIRCULATION, 1988, 77 (02) :316-327
[7]  
DEPUEY EG, 1989, J NUCL MED, V30, P441
[8]  
DEPUEY EG, 1989, J NUCL MED, V30, P548
[9]   THE INCONSISTENT PATTERN OF THALLIUM DEFECTS - A CLUE TO THE FALSE POSITIVE PERFUSION SCINTIGRAM [J].
DUNN, RF ;
WOLFF, L ;
WAGNER, S ;
BOTVINICK, EH .
AMERICAN JOURNAL OF CARDIOLOGY, 1981, 48 (02) :224-232
[10]  
EISNER R, 1988, J NUCL MED, V29, P91