Accuracy of methemoglobin measurements:: Comparison of six different commercial devices and one manual method

被引:12
作者
Dötsch, J
Demirakça, S
Hahn, D
Katz, N
Kühl, PG
Rascher, W
机构
[1] Univ Erlangen Nurnberg, Klin Kinder & Jugendl, Dept Pediat, D-91054 Erlangen, Germany
[2] Univ Giessen, Dept Pediat, Giessen, Germany
[3] Univ Giessen, Dept Clin Chem & Pathobiochem, Giessen, Germany
关键词
adverse effects; critical care; hemoglobinometry; interference; methemoglobin; methemoglobinemia; methylene blue; monitoring; nitric oxide; oximetry;
D O I
10.1097/00003246-199906000-00051
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: During nitric oxide inhalation, methemoglobinemia needs to be monitored. We compared six commercially available instruments and one manual method for methemoglobin measurements. In addition, we studied whether and to what degree methylene blue interferes with methemoglobin measurements, Design: In vitro methodologic study. Setting: Research laboratory in a university hospital. Patients: Five healthy volunteers from whom red blood cells were obtained. Interventions: Methemoglobinemia was generated in a red blood cell suspension by nitric oxide; methemoglobin was measured with six commercial instruments and one manual photometric method to calculate variation coefficients and to determine the differences between the devices. Methemoglobin was measured with and without the addition of methylene blue with two instruments. Measurements were performed immediately after the addition of methylene blue. Measurements and Main Results: All six commercially available instruments had variation coefficients of <0.1 at methemoglobin concentrations of 5%, whereas the manual photometric method did not reach a variation coefficient of <0.1 at 8% of methemoglobin. Apart from two devices that measured slightly but significantly higher methemoglobin levels, all instruments measured similar values of methemoglobin when the same samples were determined simultaneously. Higher concentrations of methylene blue (10, 40, 100 mu M) reduced substantially the apparent concentrations of methemoglobin. Interference by methylene blue was most pronounced at low methemoglobin levels. Conclusions: With some limitations, all commercial instruments that were tested performed adequately for the monitoring of methemoglobinemia, Methylene blue interferes with the methemoglobin measurements in a dose-dependent manner.
引用
收藏
页码:1191 / 1194
页数:4
相关论文
共 19 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Inhaled nitric oxide in neonatal and pediatric acute respiratory distress syndrome: Dose response, prolonged inhalation, and weaning [J].
Demirakca, S ;
Dotsch, J ;
Knothe, C ;
Magsaam, J ;
Reiter, HL ;
Bauer, J ;
Kuehl, PG .
CRITICAL CARE MEDICINE, 1996, 24 (11) :1913-1919
[3]  
DEMIRAKCA S, 1994, INTENS CARE MED, V8, P616
[4]   Extracorporeal circulation increases nitric oxide-induced methemoglobinemia in vivo and in vitro [J].
Dotsch, J ;
Demirakca, S ;
Hamm, R ;
Knothe, C ;
Bauer, J ;
Kuhl, PG ;
Rascher, W .
CRITICAL CARE MEDICINE, 1997, 25 (07) :1153-1158
[5]  
Evelyn KA, 1938, J BIOL CHEM, V126, P655
[6]   NEAR-PATIENT MEASUREMENTS OF METHEMOGLOBIN, OXYGEN-SATURATION, AND TOTAL HEMOGLOBIN - EVALUATION OF A NEW INSTRUMENT FOR ADULT AND NEONATAL INTENSIVE-CARE [J].
GONG, AK .
CRITICAL CARE MEDICINE, 1995, 23 (01) :193-201
[7]  
GRUETTER CA, 1981, J PHARMACOL EXP THER, V219, P181
[8]  
HANKS JH, 1949, P SOC EXP BIOL MED, V71, P196
[9]   INHIBITION OF VASCULAR NITRIC-OXIDE CGMP PATHWAY BY PLASMA FROM ISCHEMIC HINDLIMB OF RATS [J].
JIN, JS ;
WEBB, RC ;
DALECY, LG .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 269 (01) :H254-H261
[10]   COMPARATIVE STUDIES ON EFFECTS OF TOLUIDINE BLUE AND METHYLENE-BLUE ON REDUCTION OF FERRIHEMOGLOBIN IN MAN AND DOG [J].
KIESE, M ;
WEGER, N ;
ZIERER, A ;
LORCHER, W .
EUROPEAN JOURNAL OF CLINICAL PHARMACOLOGY, 1972, 4 (02) :115-+