Everted leaflet of a bovine pericardial aortic valve

被引:2
作者
Orihashi, Kazumasa [1 ]
Kurosaki, Tatsuya [1 ]
Sueda, Taijiro [1 ]
机构
[1] Hiroshima Univ, Dept Cardiovasc Surg, Minami Ku, Hiroshima 7348551, Japan
关键词
Bioprosthetic valve; Aortic valve; Echocardiography;
D O I
10.1510/icvts.2009.231951
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report on an everted leaflet of a Carpentier-Edwards Perimount (CEP) valve in an aortic position. A 71-year-old female patient with aortic regurgitation underwent valve replacement with a 21-mm CEP valve. After aortic declamping, transesophageal echocardiography (TEE) revealed significant transvalvular regurgitation due to an immobilized non-coronary leaflet. Upon re-aortotomy, we found an everted leaflet without any structural damage to the valve. It was readily corrected and eversion did not recur spontaneously. The valve was not replaced because it had no intrinsic problems, eversion was unlikely to be induced by cardiac output, and there was no evidence that reimplantation would avoid further problems. Her postoperative course was uneventful without recurrence of aortic regurgitation during three months of follow-up after surgery. On TEE measurements, allocation of stent posts was slightly uneven. Since leaflet eversion was re-created by distortion of the stent in an ex-vivo model, a mild imbalance of sutures might increase the risk of this event. TEE was beneficial for detection, decision-making, and avoidance of an early reoperation. This case indicates that reimplantation or replacement of the valve is not always necessary for an everted valve leaflet. Although the patient has had an uneventful course thus far, careful follow-up is mandatory. (C) 2010 Published by European Association for Cardio-Thoracic Surgery. All rights reserved.
引用
收藏
页码:1059 / 1060
页数:2
相关论文
共 3 条
[1]  
Buklas Dimitrios, 2008, Interact Cardiovasc Thorac Surg, V7, P1110, DOI 10.1510/icvts.2008.180448
[2]   Immediate structural valve deterioration of 27-mm Carpentier-Edwards aortic pericardial bioprosthesis [J].
Fleisher, AG ;
Lafaro, RJ ;
Moggio, RA .
ANNALS OF THORACIC SURGERY, 2004, 77 (04) :1443-1445
[3]   Severe, Transient Aortic Insufficiency After Bovine Pericardial Aortic Valve Replacement: A Cautionary Note [J].
Salm, Thomas J. Vander ;
Toran, Ann J. ;
Lewis, Wilfred ;
O'Connor, Johanna .
ANNALS OF THORACIC SURGERY, 2009, 88 (02) :672-674