Minimizing contrast agent dose during intraarterial gadolinium-enhanced MR angiography: In vitro assessment

被引:28
作者
Hwang, KP [1 ]
Green, JD [1 ]
Li, DB [1 ]
Simonetti, OP [1 ]
Resnick, SA [1 ]
Finn, JP [1 ]
Duerk, JL [1 ]
Omary, RA [1 ]
机构
[1] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
关键词
D O I
10.1002/jmri.10027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To minimize contrast agent dosage for intra-arterial (LA) contrast-enhanced magnetic resonance angiography (CE-MRA) by examining the effects of encoding order (elliptical vs. sequential) and injection duration (100% to 30% of the acquisition time). Materials and Methods: Catheter-based IA gadolinium (Gd) injections were performed in an arterial flow phantom. Blood flow rates, injection rates, and injection durations were systematically varied. Signal-to-noise (SNR) measurements were obtained in the aorta, renal artery, and common iliac artery. Results: No significant SNR losses were observed for any of the vessels with 75% injection duration, or for the aorta and iliac artery with 50% injection duration. Excellent images of all vessels were obtained at 5M/b injection duration. There was no significant SNR difference between encoding schemes. Conclusion: Contrast agent dosage can be substantially reduced without loss of SNR by limiting injection to part of the imaging acquisition time.
引用
收藏
页码:55 / 61
页数:7
相关论文
共 20 条
[1]   MR-guided balloon angioplasty: In vitro demonstration of the potential of MRI for guiding, monitoring, and evaluating endovascular interventions [J].
Bakker, CJ ;
Smits, HF ;
Bos, C ;
van der Weide, R ;
Zuiderveld, KJ ;
van Vaals, JJ ;
Hurtak, WF ;
Viergever, MA ;
Mall, WP .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :245-250
[2]  
Bartels LW, 2000, J MAGN RESON IMAGING, V12, P599, DOI 10.1002/1522-2586(200010)12:4<599::AID-JMRI13>3.3.CO
[3]  
2-Z
[4]  
Bos C, 2000, MAGN RESON MED, V44, P575, DOI 10.1002/1522-2594(200010)44:4<575::AID-MRM11>3.0.CO
[5]  
2-S
[6]  
Buecker A, 2000, J MAGN RESON IMAGING, V12, P616, DOI 10.1002/1522-2586(200010)12:4<616::AID-JMRI15>3.0.CO
[7]  
2-F
[8]  
Fain SB, 1999, MAGNET RESON MED, V42, P1106, DOI 10.1002/(SICI)1522-2594(199912)42:6<1106::AID-MRM15>3.0.CO
[9]  
2-Q
[10]   Determination of optimal injection parameters for intraarterial gadolinium-enhanced MR angiography [J].
Frayne, R ;
Omary, RA ;
Unal, O ;
Strother, CM .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (10) :1277-1284