Utility of whole blood hemostatometry using the clot signature analyzer® for assessment of hemostasis in cardiac surgery

被引:11
作者
Faraday, N
Guallar, E
Sera, VA
Bolton, ED
Scharpf, RB
Cartarius, AM
Emery, K
Concord, J
Kickler, TS
机构
[1] Johns Hopkins Med Inst, Welch Ctr Prevent Epidemiol & Clin Res, Baltimore, MD 21205 USA
[2] Johns Hopkins Med Inst, Dept Anesthesiol & Crit Care Med, Baltimore, MD 21205 USA
[3] Johns Hopkins Med Inst, Dept Epidemiol, Baltimore, MD 21205 USA
[4] Johns Hopkins Med Inst, Dept Clin Res, Baltimore, MD 21205 USA
[5] Johns Hopkins Med Inst, Dept Pathol, Baltimore, MD 21205 USA
关键词
D O I
10.1097/00000542-200205000-00014
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: A hemostatic monitor capable of rapid, accurate detection of clinical coagulopathy within the operating room could improve management of bleeding after cardiopulmonary bypass (CPB). The Clot Signature Analyzer(R) is a hemostatometer that measures global hemostasis in whole blood. The authors hypothesized that point-of-care hemostatometry could detect a clinical coagulopathic state in cardiac surgical patients. Methods: Fifty-seven adult patients scheduled for a variety of elective cardiac surgical procedures were studied. Anesthesia, CPB, heparin anticoagulation, prolamine reversal, and transfusion for post-CPB bleeding were all managed by standardized protocol. Clinical coagulopathy was defined by the need for platelet or fresh frozen plasma transfusion. The Clot Signature Analyzer(R) collagen-induced thrombus formation (CITF) assay measured platelet-mediated hemostasis in vitro. The activated clotting time, platelet count, prothrombin time, activated partial thromboplastin time, and fibrinogen concentration were also measured. Results: The postprotamine CITF was greater in patients who required hemostatic transfusion than in those who did not (17.6 +/- 8.0 min vs. 10.5 +/- 5.7 min, respectively; P < 0.01). Postprotamine CITF values were highly correlated with platelet and fresh frozen plasma transfusion (Spearman r = 0.50, P < 0.001 and r = 0.40, P < 0.005, respectively). Receiver operator characteristic curves showed a highly significant relation between the postprotamine CITF and intraoperative platelet and fresh frozen plasma transfusion (area under the curve, 0.78-0.81, P < 0.005) with 60-80% sensitivity, specificity, positive and negative predictive values at cutoffs of 12-14 min. Logistic regression demonstrated that the CITY was independently predictive of post-CPB hemostatic transfusion, but standard hemostatic assays were not. Conclusions: The Clot Signature Analyzer(R) CITF detects a clinical coagulopathic state after CPB and is independently predictive of the need for hemostatic transfusion. Hemostatometry has potential utility for monitoring hemostasis in cardiac surgery.
引用
收藏
页码:1115 / 1122
页数:8
相关论文
共 28 条
[1]   Hemostasis management by use of Hepcon/HMS™:: Increased bleeding without increased need for blood transfusion [J].
Beholz, S ;
Grubitzsch, H ;
Bergmann, B ;
Wollert, HG ;
Eckel, L .
THORACIC AND CARDIOVASCULAR SURGEON, 1999, 47 (05) :322-327
[2]   THE EFFECT OF AN INTRAOPERATIVE TREATMENT ALGORITHM ON PHYSICIANS TRANSFUSION PRACTICE IN CARDIAC-SURGERY [J].
DESPOTIS, GJ ;
GRISHABER, JE ;
GOODNOUGH, LT .
TRANSFUSION, 1994, 34 (04) :290-296
[3]   Evaluation of a new point-of-care test that measures PAF-mediated acceleration of coagulation in cardiac surgical patients [J].
Despotis, GJ ;
Levine, V ;
Filos, KS ;
Santoro, SA ;
Joist, JH ;
Spitznagel, E ;
Goodnough, LT .
ANESTHESIOLOGY, 1996, 85 (06) :1311-1323
[4]   Factors associated with excessive postoperative blood loss and hemostatic transfusion requirements: A multivariate analysis in cardiac surgical patients [J].
Despotis, GJ ;
Filos, KS ;
Zoys, TN ;
Hogue, CW ;
Spitznagel, E ;
Lappas, DG .
ANESTHESIA AND ANALGESIA, 1996, 82 (01) :13-21
[5]   Does the platelet-activated clotting test (HemoSTATUS(R)) predict blood loss and platelet dysfunction associated with cardiopulmonary bypass? [J].
Ereth, MH ;
Nuttall, GA ;
Klindworth, JT ;
MacVeigh, I ;
Santrach, PJ ;
Orszulak, TA ;
Harmsen, WS ;
Oliver, WC .
ANESTHESIA AND ANALGESIA, 1997, 85 (02) :259-264
[6]   The platelet thrombin receptor and postoperative bleeding [J].
Ferraris, VA ;
Ferraris, SP ;
Singh, A ;
Fuhr, W ;
Koppel, D ;
McKenna, D ;
Rodriguez, E ;
Reich, H .
ANNALS OF THORACIC SURGERY, 1998, 65 (02) :352-358
[7]   HEMOSTATOMETER - A NEW INVITRO TECHNIQUE FOR ASSESSING HEMOSTATIC ACTIVITY OF BLOOD [J].
GOROG, P ;
AHMED, A .
THROMBOSIS RESEARCH, 1984, 34 (04) :341-357
[8]   TRANSIENT EFFECT OF ASPIRIN ON COLLAGEN-INDUCED PLATELET ACCUMULATION [J].
GOROG, P ;
KAKKAR, VV .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1986, 86 (03) :311-316
[9]   COAGULATION OF FLOWING NATIVE BLOOD - ADVANTAGES OVER STAGNANT (TUBE) CLOTTING TESTS [J].
GOROG, P ;
KOVACS, IB .
THROMBOSIS RESEARCH, 1991, 64 (05) :611-619
[10]  
GRAM J, 1990, THROMB HAEMOSTASIS, V63, P241