Removal of Collagen Nerve Conduits (NeuraGen) after Unsuccessful Implantation: Focus on Histological Findings

被引:23
作者
Liodaki, Eirini [1 ]
Bos, Inge [2 ]
Lohmeyer, Joern A. [3 ]
Senyaman, Oezge [1 ]
Mauss, Karl Ludwig [1 ]
Siemers, Frank [1 ]
Mailaender, Peter [1 ]
Stang, Felix [1 ]
机构
[1] Univ Hosp Schleswig Holstein, Dept Plast Reconstruct & Hand Surg, Lubeck, Germany
[2] Univ Schleswig Holstein, Inst Pathol, Lubeck, Germany
[3] Tech Univ Munich, Univ Hosp Rechts Isar, Dept Plast Reconstruct & Hand Surg, D-80290 Munich, Germany
关键词
NeuraGen; collagen nerve conduit explantations; histological characteristics; REPAIR; TUBES; RECONSTRUCTION; REGENERATION; GRAFT; GUIDE;
D O I
10.1055/s-0033-1348033
中图分类号
R61 [外科手术学];
学科分类号
摘要
Nerve conduits are nonneural, hollow tubular structures designed to bridge the gap of a sectioned nerve, to protect the nerve from scar formation, and to guide the regenerating fibers into the distal nerve stump. In the 8-year experience of our department, four patients aged 14 to 50 years had an unsuccessful implantation of a nerve conduit (NeuraGen, Integra, Plainsboro, NJ). In these four patients, the collagen tubes were replaced by an autogenous nerve graft. The histological specimens showed characteristic histological findings of a scar neuroma without any signs of foreign body reaction in three cases and with minimal foreign body reaction in one case. The collagen nerve tube was completely resorbed in all cases after a time period of 6 to 17 months and could not be detected marco- or microscopically. To our knowledge, this is the first report in the English and German literature describing the histological characteristics of explanted collagen nerve tubes in humans.
引用
收藏
页码:517 / 521
页数:5
相关论文
共 22 条
[1]   Technical Use of Synthetic Conduits for Nerve Repair [J].
Agnew, Sonya P. ;
Dumanian, Gregory A. .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2010, 35A (05) :838-841
[2]  
ARCHIBALD SJ, 1995, J NEUROSCI, V15, P4109
[3]   Peripheral nerve regeneration through guidance tubes [J].
Belkas, JS ;
Shoichett, MS ;
Midha, R .
NEUROLOGICAL RESEARCH, 2004, 26 (02) :151-160
[4]   Early clinical experience with collagen nerve tubes in digital nerve repair [J].
Bushnell, Brandon D. ;
McWilliams, Andrew D. ;
Whitener, George B. ;
Messer, Terry M. .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2008, 33A (07) :1081-1087
[5]   Nerve Conduits for Nerve Repair or Reconstruction [J].
Deal, D. Nicole ;
Griffin, Justin W. ;
Hogan, MaCalus V. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2012, 20 (02) :63-68
[6]   LONG-TERM IN VIVO REGENERATION OF PERIPHERAL NERVES THROUGH BIOENGINEERED NERVE GRAFTS [J].
di Summa, P. G. ;
Kalbermatten, D. F. ;
Pralong, E. ;
Raffoul, W. ;
Kingham, P. J. ;
Terenghi, G. .
NEUROSCIENCE, 2011, 181 :278-291
[7]   A Bioabsorbable collagen nerve cuff (NeuraGen) for repair of lingual and inferior alveolar nerve injuries: A case series [J].
Farole, Anthony ;
Jamal, Basem T. .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2008, 66 (10) :2058-2062
[8]   Facial nerve repair accomplished by the interposition of a collagen nerve guide [J].
Kitahara, AK ;
Nishimura, Y ;
Shimizu, Y ;
Endo, K .
JOURNAL OF NEUROSURGERY, 2000, 93 (01) :113-120
[9]  
LI S-T, 1992, Clinical Materials, V9, P195, DOI 10.1016/0267-6605(92)90100-8
[10]  
Lohmeyer J, 2007, CHIRURG, V78, P142, DOI 10.1007/s00104-006-1269-1