Expiratory asynchrony in proportional assist ventilation

被引:22
作者
Du, HL
Ohtsuji, M
Shigeta, M
Chao, DC
Sasaki, K
Usuda, Y
Yamada, Y [1 ]
机构
[1] Yokohama City Univ, Sch Med, Dept Anesthesiol, Yokohama, Kanagawa 232, Japan
[2] Newport Med Instruments Inc, Dept Clin Res, Newport Beach, CA 92658 USA
[3] Tokyo Univ Hosp, Dept Anesthesiol, Tokyo 113, Japan
[4] Emergency Life Saving Technique Acad Tokyo, Tokyo, Japan
[5] Barlow Resp Resp Ctr, Los Angeles, CA USA
关键词
mechanical ventilation; patient-ventilator synchrony; proportional assist ventilation; computer simulation;
D O I
10.1164/ajrccm.165.7.2103049
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
One of the proposed advantages of proportional assist ventilation (PAV) has been the automatic synchrony between the end of the patient's inspiratory effort and the ventilator cycle (i.e., expiratory synchrony). However, recent clinical studies have shown a prolonged ventilator inspiratory time or even a "runaway" phenomenon with the normal use of PAV. We hypothesize that control-system delay may account for this, because in reality there is always some degree of delays between control-system's input and output in all ventilators. Computer simulation study to date has not taken into account the potential effect of control-system delay on expiratory synchrony. We therefore created a computer model in which the parameter of control-system delay time was introduced. We found that significant expiratory asynchrony may occur with this more realistic model of PAV. The ventilator flow termination may fall behind the completion of the patient inspiration by as long as 0.33 seconds under the selected simulation conditions. The inspiratory termination delay time is in proportion to the control-system delay time, the respiratory time constant, and the assist gain settings. In conclusion, this model indicates that due to the unavoidable control-system delay in the ventilators, expiratory asynchrony may be an inherent shortcoming associated with PAV.
引用
收藏
页码:972 / 977
页数:6
相关论文
共 24 条
[1]   RESPIRATORY MECHANICS DURING HALOTHANE ANESTHESIA AND ANESTHESIA-PARALYSIS IN HUMANS [J].
BEHRAKIS, PK ;
HIGGS, BD ;
BAYDUR, A ;
ZIN, WA ;
MILICEMILI, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1983, 55 (04) :1085-1092
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   PARTITIONING OF RESPIRATORY FLOW RESISTANCE IN MAN [J].
FERRIS, BG ;
MEAD, J ;
OPIE, LH .
JOURNAL OF APPLIED PHYSIOLOGY, 1964, 19 (04) :653-&
[4]   Response of ventilator-dependent patients to different levels of pressure support and proportional assist [J].
Giannouli, E ;
Webster, K ;
Roberts, D ;
Younes, M .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1999, 159 (06) :1716-1725
[5]   Compensation for increase in respiratory workload during mechanical ventilation - Pressure-support versus proportional-assist ventilation [J].
Grasso, S ;
Puntillo, F ;
Mascia, L ;
Ancona, G ;
Fiore, T ;
Bruno, F ;
Slutsky, AS ;
Ranieri, VM .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2000, 161 (03) :819-826
[6]   VARIABILITY OF PATIENT-VENTILATOR INTERACTION WITH PRESSURE SUPPORT VENTILATION IN PATIENTS WITH CHRONIC OBSTRUCTIVE PULMONARY-DISEASE [J].
JUBRAN, A ;
VANDEGRAAFF, WB ;
TOBIN, MJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1995, 152 (01) :129-136
[7]  
MACINTYRE N R, 1988, Respiratory Care, V33, P121
[8]  
MILIEMILL J, 1986, HDB PHYSL, P631
[9]   Respiratory comfort and breathing pattern during volume proportional assist ventilation and pressure support ventilation:: A study on volunteers with artificially reduced compliance [J].
Mols, G ;
von Ungern-Sternberg, B ;
Rohr, E ;
Haberthür, C ;
Geiger, K ;
Guttmann, J .
CRITICAL CARE MEDICINE, 2000, 28 (06) :1940-1946
[10]  
MURCIANO D, 1982, AM REV RESPIR DIS, V126, P837