Determination of major and trace elements in biological materials by microwave induced plasma optical emission spectrometry (MIP-OES) following tetramethylammonium hydroxide (TMAH) solubilization

被引:46
作者
Matusiewicz, H [1 ]
Golik, B [1 ]
机构
[1] Politechnika Poznanska, Dept Analyt Chem, PL-60965 Poznan, Poland
关键词
multi-element determination; biological reference materials; microwave induced plasma optical emission spectrometry; tetramethy-lammonium; hydroxide solubilization;
D O I
10.1016/j.microc.2003.10.007
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple, alternative method to acid digestion is presented for the preparation of biological materials for major and trace elements by microwave induced plasma optical emission spectrometry (MIP-OES). Standard reference materials were solubilized using a methanolic solution of tetramethylammonium hydroxide (TMAH) by the application of ultrasonic agitation. Following dilution with water suspensions were pumped to a V-groove Babington-type nebulizer then into argon plasma. Limits of detection and precision are reported for major (Na, K, Ca, Mg) and trace (Cu, Fe, Sr, Zn) elements. Standard addition procedures were used to minimize possible matrix interferences. The method offers relatively good precision (R.S.D. ranged from 6 to 12%). Measured concentrations are in satisfactory agreement with certified values for the biological reference materials: TORT-1 (Lobster hepatopancreas), IAEA-153 (Milk powder) and NIST 1577 (Bovine liver). The simple procedure could be useful as a routine analysis of biological materials. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:23 / 29
页数:7
相关论文
共 32 条
[1]   Elimination of the carbon build-up effect in the ETAAS analysis of low diluted serum samples using an in situ wet digestion with tetramethylammonium hydroxide.: Application to aluminium determination [J].
Almeida, AA ;
Lima, JLFC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (08) :1019-1024
[2]   The development of microplasmas for spectrochemical analysis [J].
Broekaert, JAC .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 374 (02) :182-187
[3]  
BROEKAERT JAC, 2000, ENCY ANAL CHEM, P4669
[4]  
Brunori C, 1999, ANN CHIM-ROME, V89, P873
[5]   A review of microwave plasma sources in atomic emission spectrometry: Literature from 1985 to the present [J].
Croslyn, AE ;
Smith, BW ;
Winefordner, JD .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1997, 27 (03) :199-255
[6]  
Dombovári J, 2001, ATOM SPECTROSC, V22, P331
[7]   Determination of As, Cd, Pb and Se in DORM-1 dogfish muscle reference material using alkaline solubilization and electrothermal atomic absorption spectrometry with Ir+Rh as permanent modifiers or Pd+Mg in solution [J].
Giacomelli, MBO ;
Lima, MC ;
Stupp, V ;
de Carvalho, RM ;
da Silva, JBB ;
Barrera, PB .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (12) :2151-2157
[8]   Development and investigation of microwave plasma techniques in analytical atomic spectrometry [J].
Jin, QH ;
Duan, YX ;
Olivares, JA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (02) :131-161
[9]   Determination of trace elements in biological materials using tetramethylammonium hydroxide for sample preparation [J].
Martins, P ;
Pozebon, D ;
Dressler, VL ;
Kemieciki, GA .
ANALYTICA CHIMICA ACTA, 2002, 470 (02) :195-204
[10]  
MATOUSEK JP, 1984, PROG ANAL ATOM SPECT, V7, P275