Volume-based dose optimization in brachytherapy

被引:12
作者
Kneschaurek, P [1 ]
Schiessl, W [1 ]
Wehrmann, R [1 ]
机构
[1] Tech Univ Munich, Klin Strahlentherapie & Radiol Onkol, Klinikum Rechts Isar, D-81675 Munich, Germany
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1999年 / 45卷 / 03期
关键词
brachytherapy; optimization; linear programming; simplex algorithm;
D O I
10.1016/S0360-3016(99)00224-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: We address the question of how to optimize the dwell time distribution in brachytherapy with a stepping source if a minimal tumor dose is prescribed within the planning target volume (PTV), Methods and Materials: For a given PTV, reference points inside and at the surface of the PTV are generated and dose constraints are prescribed. The dose at these reference points can be calculated if the positions of the sources are known. We determine a set of dwell times such that the dose constraints are fulfilled, and at the same time, the total irradiation time is minimized. The simplex algorithm allows us to find a solution (if any exists) for this problem. Results: The performance of this method has been tested for a geometrically simple PTV, This method gives better results than conventionally used algorithms for dwell time optimization. Conclusion: The method described in this paper allows a volume-oriented optimization for brachytherapy dose distribution. The algorithm guarantees finding a dwell time distribution which fulfills the prescribed dose constraints, if any solution exists. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:811 / 815
页数:5
相关论文
共 13 条
[1]  
[Anonymous], RADIUM DOSAGE MANCHE
[2]   A conformal index (COIN) to evaluate implant quality and dose specification in brachytherapy [J].
Baltas, D ;
Kolotas, C ;
Geramani, K ;
Mould, RF ;
Ioannidis, G ;
Kekchidi, M ;
Zamboglou, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (02) :515-524
[3]  
Busch M., 1984, Proceedings of the Eighth International Conference on the Use of Computers in Radiation Therapy (Cat. No. 84CH2048-7), P399
[4]  
Dantzig G. B., 1963, LINEAR PROGRAMMING E
[5]  
KNESCHAUREK P, 1997, NEW DEV INTERSTITIAL, P30
[6]  
KNESCHAUREK P, 1992, ART TUMOR RESPONSE M, P417
[7]   THE EVALUATION OF OPTIMIZED IMPLANTS FOR IDEALIZED IMPLANT GEOMETRIES [J].
LOW, DA ;
WILLIAMSON, JF .
MEDICAL PHYSICS, 1995, 22 (09) :1477-1485
[8]   PARIS SYSTEM IN INTERSTITIAL RADIATION-THERAPY [J].
PIERQUIN, B ;
DUTREIX, A ;
PAINE, CH ;
CHASSAGNE, D ;
MARINELLO, G ;
ASH, D .
ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1978, 17 (01) :33-48
[9]  
Press W. H., 1994, NUMERICAL RECIPES C
[10]  
QUIMBY E, 1970, PHYSICAL FDN RADIOLO