Validation of 4D-MSPECT and QGS for quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPET:: comparison with cardiac magnetic resonance imaging

被引:54
作者
Lipke, CSA
Kühl, HP
Nowak, B
Kaiser, HJ
Reinartz, P
Koch, KC
Buell, U
Schaefer, WM
机构
[1] Rhein Westfal TH Aachen, Univ Hosp, Dept Nucl Med, D-52074 Aachen, Germany
[2] Rhein Westfal TH Aachen, Univ Hosp, Med Clin Cardiol 1, D-52074 Aachen, Germany
关键词
gated Tc-99m-MIBI SPET; 4D-MSPECT; QGS; cardiovascular MRI; left ventricular volumes;
D O I
10.1007/s00259-003-1411-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The main aim of this study was to validate the accuracy of 4D-MSPECT in the assessment of left ventricular (LV) end-diastolic/end-systolic volumes (EDV, ESV) and ejection fraction (LVEF) from gated technetium-99m methoxyisobutylisonitrile single-photon emission tomography (Tc-99m-MIBI SPET), using cardiac magnetic resonance imaging (cMRI) as the reference method. By further comparing 4D-MSPECT and QGS with cMRI, the software-specific characteristics were analysed to elucidate clinical applicability. Fifty-four patients with suspected or proven coronary artery disease (CAD) were examined with gated Tc-99m-MIBI SPET (8 gates/cardiac cycle) about 60 min after tracer injection at rest. LV EDV, ESV and LVEF were calculated from gated Tc-99m-MIBI SPET using 4D-MSPECT and QGS. On the same day, cMRI (20 gates/cardiac cycle) was performed, with LV EDV, ESV and LVEF calculated using Simpson's rule. Both algorithms worked with all data sets. Correlation between the results of gated Tc-99m-MIBI SPET and cMRI was high for EDV [R=0.89 (4D-MSPECT), R=0.92 (QGS)], ESV [R=0.96 (4D-MSPECT), R=0.96 (QGS)] and LVEF [R=0.89 (4D-MSPECT), R=0.90 (QGS)]. In contrast to ESV, EDV was significantly underestimated by 4D-MSPECT and QGS compared to cMRI [130+/-45 ml (4D-MSPECT), 122+/-41 ml (QGS), 139+/-36 ml (cMRI)]. For LVEF, 4D-MSPECT and cMRI revealed no significant differences, whereas QGS yielded significantly lower values than cMRI [57.5%+/-13.7% (4D-MSPECT), 52.2%+/-12.4% (QGS), 60.0%+/-15.8% (cMRI)]. In conclusion, agreement between gated Tc-99m-MIBI SPET and cMRI is good across a wide range of clinically relevant LV volume and LVEF values assessed by 4D-MSPECT and QGS. However, algorithm-varying underestimation of LVEF should be accounted for in the clinical context and limits interchangeable use of software.
引用
收藏
页码:482 / 490
页数:9
相关论文
共 34 条
[1]   Myocardial viability testing and impact of revascularization on prognosis in patients with coronary artery disease and left ventricular dysfunction: A meta-analysis [J].
Allman, KC ;
Shaw, LJ ;
Hachamovitch, R ;
Udelson, JE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (07) :1151-1158
[2]   MR evaluation of ventricular function: True fast imaging with steady-state precession versus fast low-angle shot cine MR imaging: Feasibility study [J].
Barkhausen, J ;
Ruehm, SG ;
Goyen, M ;
Buck, T ;
Laub, G ;
Debatin, JF .
RADIOLOGY, 2001, 219 (01) :264-269
[3]   Left ventricular function: Correlation of quantitative gated SPECT and MR imaging over a wide range of values [J].
Bavelaar-Croon, CDL ;
Kayser, HWM ;
van der Wall, EE ;
de Roos, A ;
Dibbets-Schneider, P ;
Pauwels, EKJ ;
Germano, G ;
Atsma, DE .
RADIOLOGY, 2000, 217 (02) :572-575
[4]   Comparison of gated single-photon emission computed tomography with magnetic resonance imaging for evaluation of left ventricular function in ischemic cardiomyopathy [J].
Bax, JJ ;
Lamb, H ;
Dibbets, P ;
Pelikan, H ;
Boersma, E ;
Viergever, EP ;
Germano, G ;
Vliegen, HW ;
de Roos, A ;
Pauwels, EKJ ;
Van der Wall, EE .
AMERICAN JOURNAL OF CARDIOLOGY, 2000, 86 (12) :1299-1305
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]  
CHUGH A, 2001, J NUCL CARDIOL, V8, pS19
[7]   Myocardial viability as a determinant of the ejection fraction response to carvedilol in patients with heart failure (CHRISTMAS trial): randomised controlled trial [J].
Cleland, JGF ;
Pennell, DJ ;
Ray, SG ;
Coats, AJ ;
Macfarlane, PW ;
Murray, GD ;
Mule, JD ;
Vered, Z ;
Lahiri, A .
LANCET, 2003, 362 (9377) :14-21
[8]  
DEPUEY E, 1995, J NUCL MED, V36, P952
[9]   QUANTIFICATION OF THE LEFT-VENTRICULAR VOLUMES AND FUNCTION WITH CINE MR-IMAGING - COMPARISON OF GEOMETRIC-MODELS WITH 3-DIMENSIONAL DATA [J].
DULCE, MC ;
MOSTBECK, GH ;
FRIESE, KK ;
CAPUTO, GR ;
HIGGINS, CB .
RADIOLOGY, 1993, 188 (02) :371-376
[10]  
Faber TL, 1999, J NUCL MED, V40, P650