Ultrafast two-dimensional x-ray imaging with x-ray streak cameras for laser fusion research (invited)

被引:30
作者
Shiraga, H
Miyanaga, N
Heya, M
Nakasuji, M
Aoki, Y
Azechi, H
Yamanaka, T
Mima, K
机构
[1] Institute of Laser Engineering, Osaka University, Suita, Osaka 565
关键词
D O I
10.1063/1.1147690
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Ultrafast two-dimensional x-ray imaging is required for diagnosing laser-driven inertial confinement fusion plasmas. Image sampling technique with x-ray streak cameras can meet this requirement. Multi-imaging x-ray streak camera method (MIXS) with temporal and spatial resolutions of 10 ps and 15 mu m, respectively, has been developed and successfully utilized for diagnosing the uniformity and heating process of the imploded core plasmas. The two-dimensional sampling-image x-ray streak camera method is also presented. Two types of spectroscopic applications of the MIXS have been developed recently. One is multichannel MIXS, which has three MIXS channels with various spectral responses for time-resolved two-dimensional temperature measurement of the plasmas. Another is monochromatic MIXS for temperature, density, and mixing measurement, in which monochromatic images with Bragg crystals are coupled to MIXS. (C) 1997 American Institute of Physics.
引用
收藏
页码:745 / 749
页数:5
相关论文
共 15 条
[1]   HIGH-DENSITY COMPRESSION EXPERIMENTS AT ILE, OSAKA [J].
AZECHI, H ;
JITSUNO, T ;
KANABE, T ;
KATAYAMA, M ;
MIMA, K ;
MIYANAGA, N ;
NAKAI, M ;
NAKAI, S ;
NAKAISHI, H ;
NAKATSUKA, M ;
NISHIGUCHI, A ;
NORRAYS, PA ;
SETSUHARA, Y ;
TAKAGI, M ;
YAMANAKA, M .
LASER AND PARTICLE BEAMS, 1991, 9 (02) :193-207
[2]   HIGH-SPEED GATED X-RAY-IMAGING FOR ICF TARGET EXPERIMENTS (INVITED) [J].
BRADLEY, DK ;
BELL, PM ;
KILKENNY, JD ;
HANKS, R ;
LANDEN, O ;
JAANIMAGI, PA ;
MCKENTY, PW ;
VERDON, CP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) :4813-4817
[3]   DEVELOPMENT AND CHARACTERIZATION OF A PAIR OF 30-40-PS X-RAY FRAMING CAMERAS [J].
BRADLEY, DK ;
BELL, PM ;
LANDEN, OL ;
KILKENNY, JD ;
OERTEL, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01) :716-718
[4]   X-RAY TECHNIQUE FOR CONTINUOUS TIME-RESOLVED AND SPACE-RESOLVED OBSERVATION OF HOT-SPOTS [J].
CHOI, P ;
ALIAGA, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10) :2747-2749
[5]   NOVEL TIME AND TWO-DIMENSIONAL SPACE-RESOLVED X-RAY-IMAGING TECHNIQUE [J].
DEENEY, C ;
CHOI, P .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (11) :3558-3559
[6]   4.8-KEV X-RAY BACKLIGHT FRAMING METHOD FOR OBSERVING IMAGES OF SOFT-X-RAY-DRIVEN FUSION CAPSULES [J].
KATAYAMA, M ;
SHIRAGA, H ;
NAKAI, M ;
KOBAYASHI, T ;
KATO, Y .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (03) :706-710
[7]   MULTIFRAME X-RAY-IMAGING SYSTEM FOR TEMPORALLY AND SPATIALLY RESOLVED MEASUREMENTS OF IMPLODING INERTIAL CONFINEMENT FUSION-TARGETS [J].
KATAYAMA, M ;
NAKAI, M ;
YAMANAKA, T ;
IZAWA, Y ;
NAKAI, S .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (01) :124-129
[8]   HIGH-SPEED PROXIMITY FOCUSED X-RAY CAMERAS [J].
KILKENNY, JD .
LASER AND PARTICLE BEAMS, 1991, 9 (01) :49-69
[10]   TIME-RESOLVED HIGH-ENERGY X-RAY-SPECTRA OF LASER-IRRADIATED TARGETS [J].
LEE, PHY ;
ROSEN, MD .
PHYSICAL REVIEW LETTERS, 1979, 42 (04) :236-239