5-FU loaded pHEMA drainage implants for glaucoma-filtering surgery: Device design and in vitro release kinetics

被引:16
作者
Gokce, M
Akata, RF
KiremitciGumusderelioglu, M
机构
[1] HACETTEPE UNIV,DEPT CHEM ENGN,ANKARA 06532,TURKEY
[2] GAZI UNIV,DEPT OPHTHALMOL,ANKARA,TURKEY
关键词
5-fluorouracil; pHEMA; drainage device; sustained release; glaucoma;
D O I
10.1016/0142-9612(96)83290-X
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Implantable monolithic and reservoir-like water-swellable drainage devices were developed for the subconjunctival sustained release of 5-fluorouracil (5-FU) in glaucoma-filtering surgery. A water-swellable matrix was formed of a copolymer of 2-hydroxyethyl methacrylate (HEMA) with different amounts of ethylene glycol dimethacrylate (EGDMA). Drug incorporation was done before polymerization and cross-linking. Briefly, to prepare the monolithic device the monomer-drug mixture is compression moulded into a 10 mm cylinder of 1 mm length. Furthermore, reservoir-like devices were obtained by coating the monolithic devices with a highly cross-linked polymer of HEMA (pHEMA) composition. The pHEMA devices containing 5-FU or not were well characterized by means of dynamic swelling studies, structural and thermal analysis. The release of 5-FU from these implants was studied in vitro. The rate of drug release was controlled by changing the drug loading (i.e. 10 mg or 20 mg 5-FU per device), cross-linking density of polymer matrix and type of implantable device, i.e. monolithic or reservoir-like device. While monolithic devices are releasing total releasable 5-FU during the first 10 h, reservoir-like devices prolong 5-FU release for up to 120 h. The 5-FU diffusion coefficient in swollen devices (D-s,D-s) is in the order of 10(-8) cm(2) s(-1) (approximately 10 times smaller than D-w,D-g values) and it is dependent on the cross-linking density of polymeric matrix and device load. These preliminary results suggested that 20 mg 5-FU-loaded reservoir-like devices may be a potentially effective system to deliver 5-FU into the subconjunctiva.
引用
收藏
页码:941 / 949
页数:9
相关论文
共 26 条
[1]   POLY(ETHYLENE OXIDE) (PEO) AND DIFFERENT MOLECULAR-WEIGHT PEO BLENDS MONOLITHIC DEVICES FOR DRUG RELEASE [J].
APICELLA, A ;
CAPPELLO, B ;
DELNOBILE, MA ;
LAROTONDA, MI ;
MENSITIERI, G ;
NICOLAI, L .
BIOMATERIALS, 1993, 14 (02) :83-90
[2]  
BLANDFORD DL, 1992, INVEST OPHTH VIS SCI, V33, P3430
[3]  
CRANK J, 1968, DIFFUSION POLYM, P258
[4]  
Davidson G.W.R., 1986, J. Control. Release, V3, P243, DOI [DOI 10.1016/0168-3659(86)90096-9, 10.1016/0168-3659(86)90096-9]
[5]   SUBCUTANEOUS POLYMERIC MATRIX SYSTEM P(HEMA-BGA) FOR CONTROLLED RELEASE OF AN ANTICANCER DRUG (5-FLUOROURACIL) .2. RELEASE KINETICS [J].
DENIZLI, A ;
KIREMITCI, M ;
PISKIN, E .
BIOMATERIALS, 1988, 9 (04) :363-366
[6]   SUBCUTANEOUS POLYMERIC MATRIX SYSTEM POLY(HEMA-BGA) FOR CONTROLLED RELEASE OF AN ANTICANCER DRUG (5-FLUOROURACIL) .1. SYNTHESIS AND STRUCTURE [J].
DENIZLI, A ;
KIREMITCI, M ;
PISKIN, E .
BIOMATERIALS, 1988, 9 (03) :257-262
[7]   Statistical mechanics of cross-linked polymer networks I Rubberlike elasticity [J].
Flory, PJ ;
Rehner, J .
JOURNAL OF CHEMICAL PHYSICS, 1943, 11 (11) :512-520
[8]  
GOKCE M, 1994, THESIS HACETTEPE U A
[9]  
GRESSEL MG, 1984, OPHTHALMOLOGY, V91, P3780
[10]   PRIMATE TRABECULECTOMIES WITH 5-FLUOROURACIL COLLAGEN IMPLANTS [J].
HASTY, B ;
HEUER, DK ;
MINCKLER, DS .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1990, 109 (06) :721-725