Comparing T1-weighted and T2-weighted three-point Dixon technique with conventional T1-weighted fat-saturation and short-tau inversion recovery (STIR) techniques for the study of the lumbar spine in a short-bore MRI machine

被引:46
作者
Brandao, S. [1 ]
Seixas, D. [1 ]
Ayres-Basto, M. [1 ]
Castro, S. [1 ]
Neto, J. [1 ]
Martins, C. [1 ]
Ferreira, J. C. [1 ]
Parada, F. [1 ]
机构
[1] Clin Med Arrifana Sousa, P-4560466 Penafiel, Portugal
关键词
INITIAL CLINICAL-EXPERIENCE; INHOMOGENEITY CORRECTION; PULSE SEQUENCE; WATER; SUPPRESSION; QUANTIFICATION; DECOMPOSITION; DISEASE; IMAGES;
D O I
10.1016/j.crad.2013.06.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
AIM: To compare T1-weighted (W) fatewater separation (Dixon's technique) with T1W fatsaturation (FS) and T2W Dixon with short-tau inversion recovery (STIR) images for fat suppression in a short-bore MRI machine. MATERIALS AND METHODS: Thirteen patients with lumbar disease were studied on using 1.5 T MRI machine. The protocol included T1 FS (with contrast medium administration) and/or STIR and T1W and/or T2W Dixon, for comparison. Three neuroradiologists scored the images from the two-pairs of techniques for fat suppression uniformity and lesion conspicuity. Clinical usefulness of fat-only images was also evaluated. RESULTS: Regarding uniformity of fat suppression, mean scores were 2.28 (+/- 0.49), 3.06 (+/- 0.49), 2.39 (+/- 0.49), and 3.15 (+/- 0.35) for T1W FS, T1W Dixon, STIR, and T2W Dixon sequences, respectively. For the same pulse sequences, lesion conspicuity scores were 2.78 (+/- 0.50), 2.78 (+/- 0.27), 2.76 (+/- 0.47), and 2.91 (+/- 0.4), respectively. Both T1W and T2W Dixon sequences showed more homogeneous fat-suppression when compared to T1W FS (p = 0.026) and STIR (p = 0.008) techniques, but no significant difference was found for lesion conspicuity. Mean scores for the diagnostic utility of fat-only maps were, respectively, 1.72 (+/- 0.39) and 2.48 (+/- 0.50) for T1W and T2W Dixon. CONCLUSION: Fat suppression quality was superior with Dixon when compared to the conventional sequences, but not lesion conspicuity, suggesting that both techniques are reliable for diagnosis. Dixon may be advantageous in cases where inhomogeneity artefacts are an issue. Water-only maps appear to be useful in the clinical setting. (C) 2013 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:E617 / E623
页数:7
相关论文
共 32 条
[1]
Battolla L, 1994, Radiol Med, V88, P353
[2]
T1-weighted Fat-suppressed Imaging of the Pelvis with a Dual-Echo Dixon Technique: Initial Clinical Experience [J].
Beddy, Peter ;
Rangarajan, R. Deepa ;
Kataoka, Masako ;
Moyle, Penelope ;
Graves, Martin J. ;
Sala, Evis .
RADIOLOGY, 2011, 258 (02) :583-589
[3]
Fat and Water Magnetic Resonance Imaging [J].
Bley, Thorsten A. ;
Wieben, Oliver ;
Francois, Christopher J. ;
Brittain, Jean H. ;
Reeder, Scott B. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 31 (01) :4-18
[4]
Reducing Metallic Artifacts in Postoperative Spinal Imaging: Usefulness of IDEAL Contrast-enhanced T1- and T2-weighted MR Imaging-Phantom and Clinical Studies [J].
Cha, Jang Gyu ;
Jin, Wook ;
Lee, Min Hee ;
Kim, Dong Hoon ;
Park, Jai Soung ;
Shin, Won Han ;
Yi, Boem Ha .
RADIOLOGY, 2011, 259 (03) :885-893
[5]
Colosimo Cesare, 2006, Neuroradiology, V48 Suppl 1, P18
[6]
Two-point dixon technique for water-fat signal decomposition with B-o inhomogeneity correction [J].
Coombs, BD ;
Szumowski, J ;
Coshow, W .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (06) :884-889
[7]
Pelvic imaging using a T1W fat-suppressed three-dimensional dual echo dixon technique at 3T [J].
Cornfeld, Daniel M. ;
Israel, Gary ;
McCarthy, Shirley M. ;
Weinreb, Jeffery C. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (01) :121-127
[8]
Breast cancer: Gadolinium-enhanced MR imaging with a 0.5-T open imager and three-point Dixon technique [J].
Daniel, BL ;
Butts, K ;
Glover, GH ;
Cooper, C ;
Herfkens, RJ .
RADIOLOGY, 1998, 207 (01) :183-190
[9]
Fat suppression in MR imaging: Techniques and pitfalls [J].
Delfaut, EM ;
Beltran, J ;
Johnson, G ;
Rousseau, J ;
Marchandise, X ;
Cotten, A .
RADIOGRAPHICS, 1999, 19 (02) :373-382
[10]
SIMPLE PROTON SPECTROSCOPIC IMAGING [J].
DIXON, WT .