Recent progress in the development of Terumo implantable left ventricular assist system

被引:22
作者
Nojiri, C
Kijima, T
Maekawa, J
Horiuchi, K
Kido, T
Sugiyama, T
Mori, T
Sugiura, N
Asada, T
Shimane, H
Ozaki, T
Suzuki, M
Akamatsu, T
Akutsu, T
机构
[1] Terumo Corp, R&D Ctr, Nakai, Kanagawa 2590151, Japan
[2] NTN Corp, Osaka, Japan
[3] Setsunan Univ, Osaka, Japan
关键词
D O I
10.1097/00002480-199905000-00018
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The research group of the Terumo Corporation, the NTN Corporation, and Setsunan University (T. Akamatsu) has been developing an implantable left ventricular assist system (ILVAS) featuring a centrifugal blood pump with a magnetically suspended impeller (MSCP). The impeller of the MSCP is suspended by a magnetic bearing, providing contact-free rotation of the impeller inside the pump housing. Thus the MSCP is expected to provide years of long-term durability. Ex vivo chronic sheep experiments using the extracorporeal model (Model I) demonstrated long-term durability, nonthrombogenicity, and a low hemolysis rate (plasma free Hb <6 mg/dl) for more than 2 years. The prototype implantable model (Model II; 196 mi, 400 g) was evaluated ex vivo in 2 sheep and intrathoracically implanted in a small sheep (45 kg). These experiments were terminated at 70, 79, and 17 days, respectively, because of blood leakage through the connector system within the housing of Model II. There was no thrombus formation on the retrieved pump surfaces. A new connector system was introduced to the Model II pump (modified Model II), and the pump was intrathoracically implanted in a sheep. Pump flow rate was maintained at 3-7 L/min at 1700-1800 rpm. The temperature elevation on the surfaces of the motor and the electromagnet inside the pump casing was kept less than 6 degrees C. The temperature of the tissue adjacent to the pump casing became normal 10 days postoperatively. The sheep survived for more than 5 months without any sign of mechanical failure or thromboembolic complication. In vitro real-time endurance tests of motor bearings made of stainless steel and silicone nitride have been conducted for more than 1 year without any sign of bearing wear. The next prototype system (Model III), with an implantable controller and a new MSCP with reduced input power, has been developed with a view toward a totally implantable LVAS.
引用
收藏
页码:199 / 203
页数:5
相关论文
共 5 条
[1]   CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
NAKAZEKI, T ;
ITOH, H .
ARTIFICIAL ORGANS, 1992, 16 (03) :305-308
[2]   RECENT STUDIES OF THE CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
TSUKIYA, T ;
NISHIMURA, K ;
PARK, CH ;
NAKAZEKI, T .
ARTIFICIAL ORGANS, 1995, 19 (07) :631-634
[3]  
Kono S, 1997, ASAIO J, V43, pM580
[4]  
Nishimura K, 1997, ASAIO J, V43, pM553
[5]   Can heparin immobilized surfaces maintain nonthrombogenic activity during in vivo long-term implantation? [J].
Nojiri, C ;
Kido, T ;
Sugiyama, T ;
Horiuchi, K ;
Kijima, T ;
Hagiwara, K ;
Kuribayashi, E ;
Nogawa, A ;
Ogiwara, K ;
Akutsu, T .
ASAIO JOURNAL, 1996, 42 (05) :M468-M475