Performance evaluation of a hybrid linear ion trap/orbitrap mass spectrometer

被引:530
作者
Makarov, A [1 ]
Denisov, E [1 ]
Kholomeev, A [1 ]
Baischun, W [1 ]
Lange, O [1 ]
Strupat, K [1 ]
Horning, S [1 ]
机构
[1] Thermo Electron Bremen GmbH, D-28199 Bremen, Germany
关键词
D O I
10.1021/ac0518811
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Design and performance of a novel hybrid mass spectrometer is described. It couples a linear ion trap mass spectrometer to an orbitrap mass analyzer via an rf-only trapping quadrupole with a curved axis. The latter injects pulsed ion beams into a rapidly changing electric field in the orbitrap wherein they are trapped at high kinetic energies around an inner electrode. Image current detection is subsequently performed after a stable electrostatic field is achieved. Fourier transformation of the acquired transient allows wide mass range detection with high resolving power, mass accuracy, and dynamic range. The entire instrument operates in LC/MS mode (1 spectrum/s) with nominal mass resolving power of 60 000 and uses automatic gain control to provide high-accuracy mass measurements, within 2 ppm using internal standards and within 5 ppm with external calibration. The maximum resolving power exceeds 100 000 (fwhm). Rapid, automated data-dependent capabilities enable real-time acquisition of up to three high-mass accuracy MS/MS spectra per second.
引用
收藏
页码:2113 / 2120
页数:8
相关论文
共 15 条
[1]   Design and performance of an ESI interface for selective external ion accumulation coupled to a Fourier transform ion cyclotron mass spectrometer [J].
Belov, ME ;
Nikolaev, EN ;
Anderson, GA ;
Udseth, HR ;
Conrads, TP ;
Veenstra, TD ;
Masselon, CD ;
Gorshkov, MV ;
Smith, RD .
ANALYTICAL CHEMISTRY, 2001, 73 (02) :253-261
[2]   An introduction to quadrupole-time-of-flight mass spectrometry [J].
Chernushevich, IV ;
Loboda, AV ;
Thomson, BA .
JOURNAL OF MASS SPECTROMETRY, 2001, 36 (08) :849-865
[3]   Quantitation of ion abundances in Fourier transform ion cyclotron resonance mass spectrometry [J].
Goodner, KL ;
Milgram, KE ;
Williams, KR ;
Watson, CH ;
Eyler, JR .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1998, 9 (11) :1204-1212
[4]   Interfacing the orbitrap mass analyzer to an electrospray ion source [J].
Hardman, M ;
Makarov, AA .
ANALYTICAL CHEMISTRY, 2003, 75 (07) :1699-1705
[5]   The Orbitrap: a new mass spectrometer [J].
Hu, QZ ;
Noll, RJ ;
Li, HY ;
Makarov, A ;
Hardman, M ;
Cooks, RG .
JOURNAL OF MASS SPECTROMETRY, 2005, 40 (04) :430-443
[6]   A method for the neutralization of electron space charge by positive ionization at very low gas pressures [J].
Kingdon, KH .
PHYSICAL REVIEW, 1923, 21 (04) :408-418
[7]   EXPERIMENTAL-DETERMINATION OF THE NUMBER OF TRAPPED IONS, DETECTION LIMIT, AND DYNAMIC-RANGE IN FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY [J].
LIMBACH, PA ;
GROSSHANS, PB ;
MARSHALL, AG .
ANALYTICAL CHEMISTRY, 1993, 65 (02) :135-140
[8]   Electrostatic axially harmonic orbital trapping: A high-performance technique of mass analysis [J].
Makarov, A .
ANALYTICAL CHEMISTRY, 2000, 72 (06) :1156-1162
[9]  
MAKAROV A, 2005, P 53 C AM SOC MASS S
[10]  
Morris HR, 1996, RAPID COMMUN MASS SP, V10, P889, DOI 10.1002/(SICI)1097-0231(19960610)10:8<889::AID-RCM615>3.0.CO