Local effect of titanium implant corrosion: an experimental study in rats

被引:55
作者
Olmedo, D. G. [1 ]
Duffo, G. [2 ]
Cabrini, R. L. [1 ,3 ]
Guglielmotti, M. B. [1 ]
机构
[1] Univ Buenos Aires, Sch Dent, Dept Oral Pathol, Natl Res Council CONICET, Buenos Aires, DF, Argentina
[2] Natl Res Council CONICET, Natl Atom Energy Commiss, Dept Mat, RA-1650 Buenos Aires, DF, Argentina
[3] Natl Atom Energy Commiss, Dept Radiobiol, Buenos Aires, DF, Argentina
关键词
titanium corrosion; metal ion release; pitting; macrophage; oral surgery; osteolysis;
D O I
10.1016/j.ijom.2008.05.013
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
The aim of this study was to evaluate histologically the biological effect of pitting corrosion and to contribute clinically relevant data on the permanence of titanium metal structures used in osteosynthesis in the body. Commercially pure titanium laminar implants (control) and commercially pure titanium laminar implants with pitting corrosion (experimental) were implanted in the tibiae of rats. At 14 days post-implantation the animals were killed. The tibiae were resected, fixed, radiographed and processed for embedding in methyl methacrylate. Percentage of bone-implant contact and peri-implant bone Volume were evaluated. The histological study of the titanium implants Submitted to pitting corrosion showed scarce bone-implant contact, it was only present in the areas with no pitting and/or surface alterations. There was a statistically significant lower percentage of bone-implant contact in the experimental group (6% +/- 4) than in the control group (26% +/- 6) (P < 0.001). Products of corrosion in the peri-implant bed, especially around the blood vessels and areas of bone marrow in the metal-tissue interface, were observed. The microchemical analysis Of Corrosion products revealed the presence of titanium. The adverse local effects caused by pitting corrosion Suggest that titanium plates and grids should be used with caution as permanent fixation structures.
引用
收藏
页码:1032 / 1038
页数:7
相关论文
共 37 条
[1]
Abramson S, 2004, BIOMATERIALS SCI INT, P67
[2]
ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
[3]
A MULTICENTER REPORT ON OSSEOINTEGRATED ORAL IMPLANTS [J].
ALBREKTSSON, T .
JOURNAL OF PROSTHETIC DENTISTRY, 1988, 60 (01) :75-84
[4]
Clinical and animal experiments on stress corrosion of titanium miniplates [J].
Bessho, K. ;
Iizuka, K.B.T. .
Clinical Materials, 1993, 14 (03) :223-227
[5]
BLACK J, 1988, CORROSION DEGRADATIO, P199
[6]
BORDEN FP, 1995, J BONE SURG BR B, V37, P679
[7]
Cabrini R L, 1993, Implant Dent, V2, P264
[8]
Cabrini RL, 2002, J HISTOTECHNOL, V25, P75
[9]
CORRELATION OF TISSUE REACTION TO CORROSION IN OSTEOSYNTHETIC DEVICES [J].
FRENCH, HG ;
COOK, SD ;
HADDAD, RJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (07) :817-828
[10]
GALVELE JR, 1983, TREATISE MAT SCI TEC, V1