Multi Physical-Chemical profiler for real-time in situ monitoring of trace metal speciation and master variables:: Development, validation and field applications

被引:34
作者
Tercier-Waeber, ML
Confalonieri, F
Riccardi, G
Sina, A
Nöel, S
Buffle, J
Graziottin, F
机构
[1] Univ Geneva, CABE, Dept Inorgan & Analyt Chem, CH-1211 Geneva, Switzerland
[2] Idronaut Srl, I-20047 Brugherio, MI, Italy
关键词
trace metals; speciation; microsensors; in situ voltammetric probe; remote monitoring; aquatic systems;
D O I
10.1016/j.marchem.2005.03.004
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
This paper describes the development and the application of a submersible Multi Physical-Chemical profiler (MPCP) for in situ measurements of priority trace metal pollutant speciation and master bio-physicochemical variables. It has been designed using advanced voltammetric microsensors, microprocessor, telemetry and "smart" data acquisition and transfer technology. The MPCP allows simultaneous reliable monitoring, down to 150 m over extended period of time, of several major fractions of Cu(II), Pb(II), and Cd(II): (i) the free metal ions which are known to be related to biological uptake, (ii) the dynamic metal species which are potentially also available for organisms, (iii) the total extractable metals which represent the reservoir of metal in the test medium, and (iv) the particulate and colloidal species, by subtracting (ii) of (iii), which play important role in transport properties and residence time. In addition, it measures also pressure, temperature, pH, oxygen, conductivity, salinity, redox E, turbidity and chlorophyll a. The MPCP can be controlled either by an operator on board, in automatic mode following pre-programmed parameters, or remotely from a land station when connected to a remote controller installed in a buoy profiler or on a platform. A detailed description of the system is given. A summary of the laboratory and on-field tests performed to characterize and validate the MPCP measurements is presented. Finally, examples of attended and remote in situ MPCP monitoring/profiling in estuaries and coastal sea waters are reported. The results indicate that the system is reliable and that in situ simultaneous measurements of the three distinct fractions of trace metals can be achieved down to concentrations at the ppt level using square wave anodic stripping voltammetry (SWASV). They also demonstrated the significant advantages of the MPCP in terms of spatial and temporal resolution of data and thus its usefulness for cost effective, more efficient environmental monitoring and pollution control. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:216 / 235
页数:20
相关论文
共 26 条
[1]   Gel-integrated microelectrode arrays for direct voltammetric measurements of heavy metals in natural waters and other complex media [J].
Belmont-Hébert, C ;
Tercier, ML ;
Buffle, J ;
Fiaccabrino, GC ;
de Rooij, NF ;
Koudelka-Hep, M .
ANALYTICAL CHEMISTRY, 1998, 70 (14) :2949-2956
[2]  
BUFFLE J, 2000, IUPAC SERIES ANAL PH
[3]   Field evaluation of an electrochemical probe for in situ screening of heavy metals in groundwater [J].
Herdan, J ;
Feeney, R ;
Kounaves, SP ;
Flannery, AF ;
Storment, CW ;
Kovacs, GTA ;
Darling, RB .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1998, 32 (01) :131-136
[4]   The determination of trace metals in estuarine and coastal waters using a voltammetric in situ profiling system [J].
Howell, KA ;
Achterberg, EP ;
Braungardt, CB ;
Tappin, AD ;
Turner, DR ;
Worsfold, PJ .
ANALYST, 2003, 128 (06) :734-741
[5]  
NOEL S, UNPUB ELECTROANALYSI
[6]   QUANTITATIVE ASSESSMENT OF WORLDWIDE CONTAMINATION OF AIR, WATER AND SOILS BY TRACE-METALS [J].
NRIAGU, JO ;
PACYNA, JM .
NATURE, 1988, 333 (6169) :134-139
[7]   Permeation liquid membrane for trace metal speciation in natural waters - Transport of liposoluble Cu(II) complexes [J].
Parthasarathy, N ;
Pelletier, M ;
Buffle, J .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1025 (01) :33-40
[8]   Hollow fiber based supported liquid membrane: a novel analytical system for trace metal analysis [J].
Parthasarathy, N ;
Pelletier, M ;
Buffle, J .
ANALYTICA CHIMICA ACTA, 1997, 350 (1-2) :183-195
[9]   Chemical characteristics of groundwater around two massive sulphide deposits in an area of previous mining contamination, Iberian Pyrite Belt, Spain [J].
Pauwels, H ;
Tercier-Waeber, ML ;
Arenas, M ;
Castroviejo, R ;
Deschamps, Y ;
Lassin, A ;
Graziottin, F ;
Elorza, FJ .
JOURNAL OF GEOCHEMICAL EXPLORATION, 2002, 75 (1-3) :17-41
[10]   Simultaneous determination and speciation of zinc, cadmium, lead, and copper in natural water with minimum handling and artifacts, by voltammetry on a gel-integrated microelectrode array [J].
Pei, JH ;
Tercier-Waeber, ML ;
Buffle, J .
ANALYTICAL CHEMISTRY, 2000, 72 (01) :161-171