25-Hydroxyvitamin D testing: challenging the performance of current automated immunoassays

被引:48
作者
Farrell, Chris [2 ]
Soldo, Joshua [3 ]
Williams, Paul [4 ,5 ]
Herrmann, Markus [1 ,2 ,4 ]
机构
[1] Dist Hosp Bolzano, Dept Clin Pathol, I-39100 Bolzano, Italy
[2] Laverty Pathol, Sydney, NSW, Australia
[3] DiaSorin Inc, Stillwater, MN USA
[4] Univ Sydney, Cent & Nepean Clin Sch, Sydney, NSW 2006, Australia
[5] Royal Prince Alfred Hosp, Sydney SW Pathol Serv, Sydney, NSW, Australia
关键词
C3-epimer; immunoassays; liquid chromatography-tandem mass spectrometry; 25-hydroxyvitamin D; VITAMIN-D STATUS; D ASSAYS; PATHOLOGY CONSULTATION; ACCURACY; HEALTH; D-3;
D O I
10.1515/cclm-2012-0522
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Clinical laboratories require accurate and precise 25-hydroxyvitamin D (25-OHD) immunoassays to allow comparison of patient results with published decision limits. However, some variation in performance has been found with the previous generation of automated 25-OHD immunoassays. This study assessed the performance of four recently released automated 25-OHD immunoassays against a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay. Methods: A total of 983 samples from apparently healthy adults, plus 253 samples chosen to challenge the performance of the assays, were analyzed by the latest generation of immunoassays from Abbott, DiaSorin, Roche and Siemens. LC-MS/MS analysis was performed on a random subset of 264 samples. The precision of the immunoassays was assessed over 5 days with samples ranging between 3.0 and 370 nmol/L in concentration. Results: Immunoassays showed significant differences in precision at 25-OHD concentrations of 3.0-86.5 nmol/L but all showed acceptable precision at higher concentrations. The DiaSorin assay agreed with LC-MS/MS across the measuring range of samples tested (7.7-425 nmol/L). The other assays showed generally good performance, but had some limitations when their performance was challenged with samples with low and high 25-OHD concentrations, heterophilic antibodies or high 25-OHD2 concentrations. The C3-epimer of 25-OHD was identified in 40.4% of healthy adults tested and was a source of analytical variance in immunoassays. Conclusions: The latest generation of 25-OHD immunoassays has improved performance compared to previous assays. However, some immunoassays can still give discrepant results and this is most apparent when immunoassays are evaluated with a range of samples that challenge their analytical performance.
引用
收藏
页码:1953 / 1963
页数:11
相关论文
共 31 条
[1]  
[Anonymous], 2001, Biological variation: from principles to practice
[2]   Development and comparison of three liquid chromatography-atmospheric pressure chemical ionization/mass spectrometry methods for determining vitamin D metabolites in human serum [J].
Bedner, Mary ;
Phinney, Karen W. .
JOURNAL OF CHROMATOGRAPHY A, 2012, 1240 :132-139
[3]   25-Hydroxyvitamin D: A Difficult Analyte [J].
Carter, Graham D. .
CLINICAL CHEMISTRY, 2012, 58 (03) :486-488
[4]  
Carter GD, 2011, CURR DRUG TARGETS, V12, P19
[5]   Human anti-animal interference in DiaSorin Liaison total 25(OH)-vitamin D assay: Towards the end of a strange story? [J].
Cavalier, Etienne ;
Carlisi, Agnes ;
Bekaert, Anne-Catherine ;
Rousselle, Olivier ;
Chapelle, Jean-Paul .
CLINICA CHIMICA ACTA, 2012, 413 (3-4) :527-528
[6]   Diagnosis of Asymptomatic Primary Hyperparathyroidism: Proceedings of the Third International Workshop [J].
Eastell, R. ;
Arnold, A. ;
Brandi, M. L. ;
Brown, E. M. ;
D'Amour, P. ;
Hanley, D. A. ;
Rao, D. Sudhaker ;
Rubin, M. R. ;
Goltzman, D. ;
Silverberg, S. J. ;
Marx, S. J. ;
Peacock, M. ;
Mosekilde, L. ;
Bouillon, R. ;
Lewiecki, E. M. .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2009, 94 (02) :340-350
[7]   State-of-the-Art Vitamin D Assays: A Comparison of Automated Immunoassays with Liquid Chromatography-Tandem Mass Spectrometry Methods [J].
Farrell, Christopher-John L. ;
Martin, Steven ;
McWhinney, Brett ;
Straub, Isabella ;
Williams, Paul ;
Herrmann, Markus .
CLINICAL CHEMISTRY, 2012, 58 (03) :531-542
[8]  
Grant WB, 2005, ALTERN MED REV, V10, P94
[9]   Accuracy of 6 Routine 25-Hydroxyvitamin D Assays: Influence of Vitamin D Binding Protein Concentration [J].
Heijboer, Annemieke C. ;
Blankenstein, Marinus A. ;
Kema, Ido P. ;
Buijs, Madelon M. .
CLINICAL CHEMISTRY, 2012, 58 (03) :543-548
[10]  
Holick MF., 2002, CURR OPIN ENDOCRINOL, V9, P87, DOI DOI 10.1097/00060793-200202000-00011