Optical coherence tomography for optical biopsy - Properties and demonstration of vascular pathology

被引:432
作者
Brezinski, ME
Tearney, GJ
Bouma, BE
Izatt, JA
Hee, MR
Swanson, EA
Southern, JF
Fujimoto, JG
机构
[1] MIT,DEPT ELECT ENGN & COMP SCI,CAMBRIDGE,MA 02139
[2] MASSACHUSETTS GEN HOSP,CARDIAC UNIT,BOSTON,MA 02114
[3] MASSACHUSETTS GEN HOSP,DEPT PATHOL,BOSTON,MA 02114
[4] HARVARD UNIV,SCH MED,BOSTON,MA
[5] MIT,ELECTR RES LAB,CAMBRIDGE,MA 02139
[6] MIT,LINCOLN LAB,LEXINGTON,MA 02173
关键词
atherosclerosis; lasers; imaging;
D O I
10.1161/01.CIR.93.6.1206
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Optical coherence tomography (OCT) is a recently developed medical diagnostic technology that uses back-reflected infrared light to perform in situ micron scale tomographic imaging. In this work, we investigate the ability of OCT to perform micron scale tomographic imaging of the internal microstructure of in vitro atherosclerotic plaques. Methods and Results Aorta and relevant nonvascular tissue were obtained at autopsy. Two-dimensional cross-sectional imaging of the exposed surface of the arterial segments was performed in vitro with OCT. A 1300-nm wavelength, super-luminescent diode light source was used that allows an axial spatial resolution of 20 mu m. The signal-to-noise ratio was 109 dB. Images were displayed in gray scale or false color, Imaging was performed over 1.5 mm into heavily calcified tissue, and a high contrast was noted between lipid- and water-based constituents, making OCT attractive for intracoronary imaging. The 20-mu m axial resolution of OCT allowed small structural details such as the width of intimal caps and the presence of fissures to be determined. The extent of lipid collections, which had a low backscattering intensity, also were well documented. Conclusions OCT represents a promising new technology for imaging Vascular microstructure with a level of resolution not previously achieved with the use of other imaging modalities. It does not require direct contact with the vessel wall and can be performed with a catheter integrated with a relatively inexpensive optical fiber. The high contrast among tissue constituents, high resolution, and ability to penetrate heavily calcified tissue make OCT an attractive new imaging technology for intracoronary diagnostics.
引用
收藏
页码:1206 / 1213
页数:8
相关论文
共 42 条
  • [1] ANGIOGRAPHIC PROGRESSION OF CORONARY-ARTERY DISEASE AND THE DEVELOPMENT OF MYOCARDIAL-INFARCTION
    AMBROSE, JA
    TANNENBAUM, MA
    ALEXOPOULOS, D
    HJEMDAHLMONSEN, CE
    LEAVY, J
    WEISS, M
    BORRICO, S
    GORLIN, R
    FUSTER, V
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 12 (01) : 56 - 62
  • [2] BADIMON L, 1992, AM J CARD IMAGING, V6, P278
  • [3] Benkeser P J, 1993, J Am Soc Echocardiogr, V6, P158
  • [4] HIGH-RESOLUTION OPTICAL COHERENCE TOMOGRAPHIC IMAGING USING A MODE-LOCKED TI-AL2O3 LASER SOURCE
    BOUMA, B
    TEARNEY, GJ
    BOPPART, SA
    HEE, MR
    BREZINSKI, ME
    FUJIMOTO, JG
    [J]. OPTICS LETTERS, 1995, 20 (13) : 1486 - 1488
  • [5] HIGH-RESOLUTION REFLECTOMETRY IN BIOLOGICAL TISSUES
    CLIVAZ, X
    MARQUISWEIBLE, F
    SALATHE, RP
    NOVAK, RP
    GILGEN, HH
    [J]. OPTICS LETTERS, 1992, 17 (01) : 4 - 6
  • [6] CLIVAZ X, 1994, SPIE P, P19
  • [7] DETERMINANTS OF COLLATERAL FILLING OBSERVED DURING SUDDEN CONTROLLED CORONARY-ARTERY OCCLUSION IN HUMAN-SUBJECTS
    COHEN, M
    SHERMAN, W
    RENTROP, KP
    GORLIN, R
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (02) : 297 - 303
  • [8] DAVIES MJ, 1985, BRIT HEART J, V53, P363
  • [9] ANGIOSCOPIC VERSUS ANGIOGRAPHIC DETECTION OF INTIMAL DISSECTION AND INTRACORONARY THROMBUS
    DENHEIJER, P
    FOLEY, DP
    ESCANED, J
    HILLEGE, HL
    VANDIJK, RB
    SERRUYS, PW
    LIE, KI
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (03) : 649 - 654
  • [10] THE OPTICAL-PROPERTIES OF AQUEOUS SUSPENSIONS OF INTRALIPID, A FAT EMULSION
    DRIVER, I
    FEATHER, JW
    KING, PR
    DAWSON, JB
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (12) : 1927 - 1930