A cooled CCD based neutron imaging system for low fluence neutron sources

被引:15
作者
Lanza, RC [1 ]
Shi, S [1 ]
McFarland, EW [1 ]
机构
[1] UNIV CALIF SANTA BARBARA,DEPT CHEM ENGN,SANTA BARBARA,CA 93106
基金
美国国家科学基金会;
关键词
D O I
10.1109/23.507063
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed a neutron detection system for accelerator based neutron radiography and tomography based on a combination of scintillation screen and large aperture optics and a cooled CCD camera.[1,2,3,4,5,6] The system is capable of detecting single neutron events and can therefore be considered as a possible detector for neutron scattering as well as conventional imaging. The system has a resolution of 0.1 mm or 1242 x 1152 pixels. The limit of image size is set by the light output of the scintillator, the light collection of the optical system, the size of the CCD and the desired signal to noise ratio. The lower limit on neutron flux is determined by the dark current of the chip. Equations for these limits have been derived and can be used to predict and optimize performance. The scintillation light output per incident neutron is large enough to permit the use of lens coupled systems with their increased flexibility and ease of implementation. The system can approach a quantum limited noise level, depending on the particular geometry used. For our current system, based on the use of NE-426 scintillator[7], a 1242 x 1152 pixel (27.9 x 25.9 mm) EEV 05-30 CCD operating at -50C, and using a 100 mm focal length, F/0.9 lens, the maximum size for the imaging screen is 0.7 m (based on 10 photoelectrons for a single detected neutron), and the lower limit for flux is 0.1 n/pixel/s based on this size screen and a typical dark current of 1 e/pixel/s.
引用
收藏
页码:1347 / 1351
页数:5
相关论文
共 7 条
[1]  
LANZA RC, 1990, C REC 1990 IEEE NUCL, P86
[2]  
LANZA RC, 1991, IEEE T NUC SCI, V38, P612
[3]  
MCFARLAND E, 1992, SPIE P, V1737, P101
[4]  
MCFARLAND EW, 1995, J ADV MATER-COVINA, V26, P3
[5]   SEGMENTED ARRAY RECONSTRUCTIONS FOR TOMOGRAPHIC IMAGING OF LARGE OBJECTS - THEORY AND AN APPLICATION TO NEUTRON COMPUTED-TOMOGRAPHY [J].
MCFARLAND, EW ;
DAI, HL ;
NA, H ;
LANZA, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :1275-1281
[6]  
MCFARLAND EW, 1991, INT M TOM TECHN IND
[7]  
Shi S., 1995, THESIS