Computed tomography-based attenuation correction in neurological positron emission tomography: evaluation of the effect of the X-ray tube voltage on quantitative analysis

被引:23
作者
Ay, MR [1 ]
Zaidi, H [1 ]
机构
[1] Univ Hosp Geneva, Div Nucl Med, CH-1211 Geneva 4, Switzerland
关键词
attenuation correction; brain imaging; PET/CT; quantification; tube voltage;
D O I
10.1097/01.mnm.0000203631.23407.23
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Background The advent of dual-modality positron emission tomography/computed tomography (PET/CT) imaging has revolutionized the practice of clinical oncology by improving lesion localization and facilitating treatment planning for radiation therapy. In addition, the use of CT images for CT-based attenuation correction (CTAC) allows the overall scanning time to be decreased and a noise-free attenuation map (mu map) to be created. The most common procedure requires a piecewise linear calibration curve acquired under standard imaging conditions to convert the patient's CT image from low effective CT energy into an attenuation map at 511 keV. Aim To evaluate the effect of the tube voltage on the accuracy of CTAC. Methods As different tube voltages are employed in current PET/CT scanning protocols, depending on the size of the patient and the region under study, the impact of using a single calibration curve on the accuracy of CTAC for images acquired at different tube voltages was investigated through quantitative analysis of the created mu maps, generated attenuation correction factors and reconstructed neurological PET data using anthropomorphic experimental phantom and clinical studies. Results For CT images acquired at 80 and 140 kVp, average relative differences of - 2.9% and 0.7%, respectively, from the images acquired at 120 kVp were observed for the absolute activity concentrations in five regions of the anthropomorphic striatal phantom when CT images were converted using a single calibration curve derived at 120 kVp. Likewise, average relative differences of 1.9% and - 0.6% were observed when CT images were acquired at 120 kVp and CTAC used calibration curves derived at 80 and 140 kVp, respectively. Conclusion The use of a single calibration curve acquired under standard imaging conditions does not affect, to a visible or measurable extent neurological PET images reconstructed using CTAC when CT images are acquired in different conditions.
引用
收藏
页码:339 / 346
页数:8
相关论文
共 21 条
[1]
AY MR, 2006, IN PRESS EUR J NUCL, V33
[2]
A generalized model for the conversion from CT numbers to linear attenuation coefficients [J].
Bai, CY ;
Shao, L ;
Da Silva, AJ ;
Zhao, Z .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) :1510-1515
[3]
BERGER MJ, 1998, 873597 NBSIR
[4]
BEYER T, 1994, NUCLEAR SCIENCE SYMPOSIUM & MEDICAL IMAGING CONFERENCE - 1994 IEEE CONFERENCE RECORD, VOLS 1-4, P1573
[5]
Head and neck malignancy: Is PET/CT more accurate than PET or CT alone? [J].
Branstetter, BF ;
Blodgett, TM ;
Zimmer, LA ;
Snyderman, CH ;
Johnson, JT ;
Raman, S ;
Meltzer, CC .
RADIOLOGY, 2005, 235 (02) :580-586
[6]
PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients [J].
Burger, C ;
Goerres, G ;
Schoenes, S ;
Buck, A ;
Lonn, AHR ;
von Schulthess, GK .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (07) :922-927
[7]
DETECT - Dual energy transmission estimation CT - for improved attenuation correction in SPECT and PET [J].
Guy, MJ ;
Castellano-Smith, IA ;
Flower, MA ;
Flux, GD ;
Ott, RJ ;
Visvikis, D .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :1261-1267
[8]
Hasegawa BH, 2006, QUANTITATIVE ANALYSIS IN NUCLEAR MEDICINE IMAGING, P35, DOI 10.1007/0-387-25444-7_2
[9]
KARNEL E, 2002, EUR J NUCL MED, V29, P346
[10]
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners [J].
Kinahan, PE ;
Hasegawa, BH ;
Beyer, T .
SEMINARS IN NUCLEAR MEDICINE, 2003, 33 (03) :166-179