A decision support system for suggesting ventilator settings: Retrospective evaluation in cardiac surgery patients ventilated in the ICU

被引:17
作者
Allerod, Charlotte [1 ,2 ]
Rees, Stephen E. [2 ]
Rasmussen, Bodil S. [1 ]
Karbing, Dan S. [2 ]
Kjaergaard, Soren [1 ]
Thorgaard, Per [1 ]
Andreassen, Steen [2 ]
机构
[1] Aarhus Univ, Aalborg Hosp, Dept Anaesthesia, DK-9000 Aalborg, Denmark
[2] Aalborg Univ, Ctr Model Based Med Decis Support, DK-9220 E Aalborg, Denmark
关键词
Mechanical ventilation; Decision support system; Physiological models;
D O I
10.1016/j.cmpb.2008.07.001
中图分类号
TP39 [计算机的应用];
学科分类号
081203 [计算机应用技术]; 0835 [软件工程];
摘要
Selecting appropriate ventilator settings decreases the risk of ventilator-induced lung injury. A decision support system (DSS) has been developed based on physiological models, which can advise on setting of tidal volume (Vt), respiratory frequency (f) and fraction of inspired oxygen (FiO(2)). The, aim of this study is to assess the feasibility of the DSS by comparing its advice with the values used in clinical practice. Data from 20 patients following uncomplicated coronary artery bypass grafting (CABG) with cardiopulmonary bypass was used to test the DSS. Ventilator settings suggested by the DSS were compared to the settings selected by the clinician. When compared to the clinician the DSS suggested: lowering FiO(2) (by median 7%, range 2-17%) at high SpO(2) and increasing FiO(2) (by median 2%, range 1-5%) at low SpO(2); lowering ventilation volume (by median 0.571 min(-1), range 0.2-1.11 min(-1)) at high pHa and increasing ventilation volume (by median 0.41 min(-1), range 0.1-0.91 min(-1)) at low pHa. Suggested changes in ventilation volume were such that simulated values of PIP were <= 22.9 cmH(2)O and respiratory frequency <= 18 breaths min(-1). In all cases, computer suggested values of FiO(2), Vt or f were consistent with maintaining sufficient oxygenation, normalising pH and obtaining low values of PIP. (C) 2008 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:205 / 212
页数:8
相关论文
共 16 条
[1]
*AC RESP DISTR SYN, 2000, NEW ENGL J MED, V342, P1301, DOI DOI 10.1056/NEJM200005043421801
[2]
BENUMOF JL, 1994, ANAESTHESIA, P594
[3]
East TD, 1999, J AM MED INFORM ASSN, P251
[4]
International consensus conferences in intensive care medicine, 1999, INTENS CARE MED, V25, P1444, DOI DOI 10.1007/S001340051097
[5]
Keeney R. L, 1993, DECISIONS MULTIPLE O, DOI DOI 10.1017/CBO9781139174084
[6]
Non-invasive estimation of shunt and ventilation-perfusion mismatch [J].
Kjaergaard, S ;
Rees, S ;
Malczynski, J ;
Nielsen, JA ;
Thorgaard, P ;
Toft, E ;
Andreassen, S .
INTENSIVE CARE MEDICINE, 2003, 29 (05) :727-734
[7]
A multicenter randomized trial of computer-driven protocolized weaning from mechanical ventilation [J].
Lellouche, Francois ;
Mancebo, Jordi ;
Jolliet, Philippe ;
Roeseler, Jean ;
Schortgen, Frederique ;
Dojat, Michel ;
Cabello, Belen ;
Bouadma, Lila ;
Rodriguez, Pablo ;
Maggiore, Salvatore ;
Reynaert, Marc ;
Mersmann, Stefan ;
Brochard, Laurent .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2006, 174 (08) :894-900
[8]
MOO RE, 1994, ANAESTHESIA, P1279
[10]
Oxygenation within the first 120 h following coronary artery bypass grafting.: Influence of systemic hypothermia (32°C) or normothermia (36°C) during the cardiopulmonary bypass:: a randomized clinical trial [J].
Rasmussen, BS ;
Sollid, J ;
Rees, SE ;
Kjærgaard, S ;
Murley, D ;
Toft, E .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2006, 50 (01) :64-71